A novel sidestream ultrasonic flow sensor for multiple breath washout in children.

Susanne I Fuchs1, J Sturz, S Junge, M Ballmann, M Gappa.   

Abstract

OBJECTIVE: Inert gas multiple breath washout (MBW) for measuring Lung Clearance Index using mass spectrometry and 4% sulfur hexafluoride (SF(6)) as the tracer gas has been shown to be sensitive for detecting early Cystic Fibrosis (CF) lung disease. However, mass spectrometry requires bulky equipment and is expensive to buy and maintain. A novel sidestream ultrasonic device may overcome this problem. The aims of this study were to assess the feasibility and clinical validity of measuring lung volume (functional residual capacity, FRC) and the LCI using the sidestream ultrasonic flow sensor in children and adolescents with CF in relation to spirometry and plain chest radiographs. PATIENTS AND METHODS: MBW using the sidestream ultrasonic device and conventional spirometry were performed in 26 patients with CF and 22 healthy controls.
RESULTS: In the controls (4.7-17.7 years) LCI was similar to that reported using mass spectrometry (mean (SD) 6.7 (0.5)). LCI was elevated in 77% of the CF children (6.8-18.9 years), whereas spirometry was abnormal in only 38.5%, 61.5%, and 26.9% for FEV(1), MEF(25), and FEV(1)/FVC, respectively. This was more marked in children <10 years. LCI correlated with the Crispin-Norman score, whereas FEV(1) did not.
CONCLUSIONS: Sidestream ultrasonic MBW is a valid and simple alternative to mass spectrometry for assessing ventilation homogeneity in children. (c) 2008 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2008        PMID: 18618619     DOI: 10.1002/ppul.20825

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  5 in total

1.  Lung Clearance Index (LCI) in Patients with Bronchiolitis Obliterans: A Preliminary Report and Comparison to Cystic Fibrosis Patients.

Authors:  Michal Gur; Karin Yaacoby-Bianu; Anat Ilivitzki; Ronen Bar-Yoseph; Vered Nir; Fahed Hakim; Yazeed Toukan; Lea Bentur
Journal:  Lung       Date:  2016-08-31       Impact factor: 2.584

2.  Tidal volume single breath washout of two tracer gases--a practical and promising lung function test.

Authors:  Florian Singer; Georgette Stern; Cindy Thamrin; Oliver Fuchs; Thomas Riedel; Per Gustafsson; Urs Frey; Philipp Latzin
Journal:  PLoS One       Date:  2011-03-10       Impact factor: 3.240

3.  Impact of Software Settings on Multiple-Breath Washout Outcomes.

Authors:  Selina Summermatter; Florian Singer; Philipp Latzin; Sophie Yammine
Journal:  PLoS One       Date:  2015-07-13       Impact factor: 3.240

4.  4-week daily airway clearance using oscillating positive-end expiratory pressure versus autogenic drainage in bronchiectasis patients: a randomised controlled trial.

Authors:  Galit Livnat; Naama Yaari; Nili Stein; Lea Bentur; Moneera Hanna; Maya Harel; Yochai Adir; Michal Shteinberg
Journal:  ERJ Open Res       Date:  2021-11-08

Review 5.  Quantification of Phenotypic Variability of Lung Disease in Children with Cystic Fibrosis.

Authors:  Mirjam Stahl; Eva Steinke; Marcus A Mall
Journal:  Genes (Basel)       Date:  2021-05-25       Impact factor: 4.096

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.