Radiation dose and image quality for paediatric interventional cardiology.

E Vano1, C Ubeda, F Leyton, P Miranda.   

Abstract

Radiation dose and image quality for paediatric protocols in a biplane x-ray system used for interventional cardiology have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4-20 cm of PMMA). Images from fluoroscopy (low, medium and high) and cine modes have been archived in digital imaging and communications in medicine (DICOM) format. Signal-to-noise ratio (SNR), figure of merit (FOM), contrast (CO), contrast-to-noise ratio (CNR) and high contrast spatial resolution (HCSR) have been computed from the images. Data on dose transferred to the DICOM header have been used to test the values of the dosimetric display at the interventional reference point. ESAK for fluoroscopy modes ranges from 0.15 to 36.60 microGy/frame when moving from 4 to 20 cm PMMA. For cine, these values range from 2.80 to 161.10 microGy/frame. SNR, FOM, CO, CNR and HCSR are improved for high fluoroscopy and cine modes and maintained roughly constant for the different thicknesses. Cumulative dose at the interventional reference point resulted 25-45% higher than the skin dose for the vertical C-arm (depending of the phantom thickness). ESAK and numerical image quality parameters allow the verification of the proper setting of the x-ray system. Knowing the increases in dose per frame when increasing phantom thicknesses together with the image quality parameters will help cardiologists in the good management of patient dose and allow them to select the best imaging acquisition mode during clinical procedures.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18612174     DOI: 10.1088/0031-9155/53/15/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Radiation dose optimisation for conventional imaging in infants and newborns using automatic dose management software: an application of the new 2013/59 EURATOM directive.

Authors:  L Alejo; E Corredoira; F Sánchez-Muñoz; C Huerga; Z Aza; R Plaza-Núñez; A Serrada; M Bret-Zurita; M Parrón; C Prieto-Areyano; G Garzón-Moll; R Madero; E Guibelalde
Journal:  Br J Radiol       Date:  2018-04-09       Impact factor: 3.039

2.  Quantification of tantalum sedimentation rates in liquid embolic agents.

Authors:  J R Mason; C Dodge; G Benndorf
Journal:  Interv Neuroradiol       Date:  2018-05-19       Impact factor: 1.610

3.  Measurement of gantry rotation time in modern ct.

Authors:  Atsushi Fukuda; Pei-Jan P Lin; Kosuke Matsubara; Tosiaki Miyati
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

4.  Biplane interventional pediatric system with cone-beam CT: dose and image quality characterization for the default protocols.

Authors:  Eva Corredoira; Eliseo Vañó; Luis Alejo; Carlos Ubeda; Federico Gutiérrez-Larraya; Julia Garayoa
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

5.  Assessment of Patient Dose with Special Look at Pediatrics during Cardiovascular Imaging.

Authors:  Mehnati P; Asghari Jafarabadi M; Danaee L
Journal:  J Biomed Phys Eng       Date:  2020-02-01