Literature DB >> 18609602

Monitoring glutamine in animal cell cultures using a chemiluminescence fiber optic biosensor.

M V Cattaneo1, J H Luong.   

Abstract

Together with flow injection analysis (FIA), a chemiluminescence (CL) fiber optic biosensor system has been developed for determining glutamine in animal cell cultures. Glutaminase (GAH) and glutamate oxidase (GLO) were onto separate porous aminopropyl glass beads via glutaraldehyde activation and packed to form an enzyme column. These two enzymes acted in sequence on glutamine to produce hydrogen peroxide, which was then reacted with luminol in the presence of ferricyanide to produce a light signal. An anion exchanger was introduced on-line to eliminate interfering endogenous glutamate in view of its negative charge at pH above 3.22 (isoelectric pH). Among several resins tested, the acetate form was most effective, and this type of ion exchanger also effectively adsorbed uric acid, acetaminophen, and aspartic acid.There was an excellent linear relationship between the CL response and standard glutamine concentration in the range 1 to 100 muM. A complete analysis could be performed in 2 min, including sampling and washing with a good reproducibility (+/- 4.4%). Both the bi-enzymic and ion exchange columns were useful for at least 500 analyses when the biosensor system was applied for the glutamine determination in murine hybridoma cell cultures and insect cell cultures. The values obtained compared well with those of HPLC, thus validating the applicability of the CL fiber optic system. (c) 1993 John Wiley & Sons, Inc.

Entities:  

Year:  1993        PMID: 18609602     DOI: 10.1002/bit.260410609

Source DB:  PubMed          Journal:  Biotechnol Bioeng        ISSN: 0006-3592            Impact factor:   4.530


  2 in total

1.  Determination of ammonium and L-glutamine in hybridoma cell cultures by sequential flow injection analysis.

Authors:  C Campmajó; J J Cairó; A Sanfeliu; E Martínez; S Alegret; F Gòdia
Journal:  Cytotechnology       Date:  1994       Impact factor: 2.058

2.  Simultaneous measurement of glucose and glutamine in aqueous solutions by near infrared spectroscopy.

Authors:  H Chung; M A Arnold; M Rhiel; D W Murhammer
Journal:  Appl Biochem Biotechnol       Date:  1995-02       Impact factor: 2.926

  2 in total