| Literature DB >> 18602241 |
Isabela da Costa César1, Fernando Henrique Andrade Nogueira, Gérson Antônio Pianetti.
Abstract
This paper describes the development and validation of a HPLC-UV method (210 nm) for the simultaneous quantitation of artemether and lumefantrine in fixed dose combination tablets. The method showed to be linear (r(2)>0.99), precise (R.S.D.<2.0%), accurate (recovery of 101.07% for artemether and 101.58% for lumefantrine), specific and robust. Four batches of artemether-lumefantrine tablets were assayed by the validated method. The artemether contents in the tablets varied from 98.61% to 103.35%, while lumefantrine contents were 97.92-100.48%.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18602241 DOI: 10.1016/j.jpba.2008.05.022
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935