Literature DB >> 18601605

The silent hemoglobin alpha chain variant Hb Riccarton [alpha51(CE9)Gly-->Ser] may affect HbA1c determination on the HLC-723 G7 analyzer.

Johannes M W van den Ouweland1, Henny van Daal, Corné H Klaassen, Yvonne van Aarssen, Cornelis L Harteveld, Piero C Giordano.   

Abstract

BACKGROUND: Structural hemoglobin variants can affect the accuracy of hemoglobin A1c (HbA1c) testing and represent the most common pitfall in the determination of HbA1c. We here describe the characterization of an alpha chain variant in diabetic patients as the cause of an abnormal presentation of the HbA1c fraction on the HLC-723 G7 analyzer.
METHODS: HbA1c analysis was performed using various HPLC-based HbA1c analyzers and by immunoassay. alpha-Globin mutation analysis was performed by GAP-PCR and DNA sequencing.
RESULTS: The peak partially overlapping HbA1c in the chromatogram represents the glycated fraction of the silent alpha chain variant Hb Riccarton [alpha51(CE9)Gly-->Ser]. This aberrant peak is uniquely identified by the HLC-723 instrument, as it is not observed on other HPLC-based HbA1c analyzers. Occasionally, the HLC-723 may fail to properly integrate both glycated Hb fractions, resulting in a falsely low HbA1c result. The variant was confirmed in samples from other diabetic patients with identical chromatographic patterns.
CONCLUSIONS: The silent alpha chain variant Hb Riccarton [alpha51(CE9)Gly-->Ser] leads to an abnormal chromatographic presentation on the HLC-723 analyzer with a risk of erroneous HbA1c determination. Manual validation of chromatograms to detect abnormalities caused by Hb variants is important to prevent incorrectly produced HbA1c results from being reported.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18601605     DOI: 10.1515/CCLM.2008.169

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  A rare haemoglobin variant (Hb Phnom Penh) manifesting as a falsely high haemoglobin A1c value on ion-exchange chromatography.

Authors:  Chi-Fen Chen; Yen-Kuang Tai
Journal:  Singapore Med J       Date:  2014-08       Impact factor: 1.858

2.  Hemoglobin variants detected by hemoglobin A1c (HbA1c) analysis and the effects on HbA1c measurements.

Authors:  Nadzimah Mohd Nasir; M Thevarajah; Chew Yee Yean
Journal:  Int J Diabetes Dev Ctries       Date:  2010-04

Review 3.  The long and winding road to optimal HbA1c measurement.

Authors:  Randie R Little; Curt L Rohlfing
Journal:  Clin Chim Acta       Date:  2013-01-11       Impact factor: 3.786