Literature DB >> 18599053

Prospective clinical trial of magnetic-anchor-guided endoscopic submucosal dissection for large early gastric cancer (with videos).

Takuji Gotoda1, Ichiro Oda, Katsunori Tamakawa, Hirohisa Ueda, Toshiaki Kobayashi, Tadao Kakizoe.   

Abstract

BACKGROUND: The treatment of early gastric cancer (EGC) by endoscopic submucosal dissection (ESD) has been rapidly gaining popularity in Japan. However, the procedure needs a high quality of skill. To facilitate complicated ESD by using a single working-channel gastroscope ("one-hand surgery method"), the magnetic-anchor-guided ESD (MAG-ESD) controlled by an extracorporeal electromagnet was reported to be successful in a porcine model.
OBJECTIVES: The purpose of this prospective clinical trial was to evaluate the feasibility of MAG-ESD for large EGC located on the gastric body in human beings.
DESIGN: Prospective clinical trial at a single center.
SETTING: National Cancer Center Hospital, Tokyo, Japan.
SUBJECTS: From January 2005 to May 2006, 25 patients with EGC >20 mm in diameter, located in the gastric body, and intestinal-type histology were enrolled. Patients with a cardiac pacemaker, advanced malignancy in other organs, severe cardiac and/or pulmonary diseases, and uncontrolled hypertension and/or diabetes mellitus were excluded from this study.
INTERVENTIONS: Similar to a standard ESD, the MAG-ESD procedure was performed with the patient under conscious sedation by intravenous injection of midazolam (3-5 mg) and pentazocine (15 mg). MAIN OUTCOME MEASUREMENTS: Unfavorable events and other intraoperative complications caused by the magnetic anchor or the magnetic force were recorded and evaluated. Two GI endoscopists (T.G., I.O.) assessed whether the magnetic anchor facilitated gastric ESD according to 2 criteria: "supportive" and "not supportive." The en bloc resection rate, complications, total operation time, bleeding, perforation, and recurrence rate were also evaluated. The total operation time was measured from insertion to withdrawal of the endoscope, including the retrieving of the magnetic anchor or anchors.
RESULTS: All tumors were resected en bloc, without any perforations or severe uncontrollable bleeding. All magnetic anchors were safely retrieved. Two endoscopists assessed that the MAG system was supportive in 23 patients. None of the patients experienced physiologic and mental abnormalities as a result of long-term magnetic-field exposure. During a median follow-up of 20 months (15-32 months), neither delayed adverse effects nor allergies caused by the stainless steel of the magnetic anchor were observed.
CONCLUSIONS: MAG-ESD is a feasible and safe method that allowed an excellent visualization by suitable tissue tension and facilitated gastric ESD in patients with EGC. The system should be miniaturized to make it applicable in daily clinical practice.

Entities:  

Mesh:

Year:  2008        PMID: 18599053     DOI: 10.1016/j.gie.2008.03.1127

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  51 in total

Review 1.  Endoscopic submucosal dissection--current success and future directions.

Authors:  Hironori Yamamoto
Journal:  Nat Rev Gastroenterol Hepatol       Date:  2012-06-05       Impact factor: 46.802

2.  Endoscopic mucosal resection and endoscopic submucosal dissection for early gastric cancer: Current and original devices.

Authors:  Keiichiro Kume
Journal:  World J Gastrointest Endosc       Date:  2009-10-15

3.  Utilisation of magnets to enhance gastrointestinal endoscopy.

Authors:  Imdadur Rahman; Praful Patel; Philip Boger; Mike Thomson; Nadeem Ahmad Afzal
Journal:  World J Gastrointest Endosc       Date:  2015-12-25

Review 4.  Future perspective of gastric cancer endotherapy.

Authors:  Takuji Gotoda; Chika Kusano; Fuminori Moriyasu
Journal:  Ann Transl Med       Date:  2014-03

5.  Traction using a clip-with-line is a preferred method for trainees in performing esophageal endoscopic submucosal dissection: an animal model study.

Authors:  Peng Jin; Kuang-I Fu; Yang Yu; Yu-Qi He; Zhi Wei; Xin Wang; Qiang Cai; Jian-Qiu Sheng
Journal:  Therap Adv Gastroenterol       Date:  2017-03-15       Impact factor: 4.409

6.  The Efficacy of an Endoscopic Grasp-and-Traction Device for Gastric Endoscopic Submucosal Dissection: An Ex Vivo Comparative Study (with Video).

Authors:  Dirk W Schölvinck; Osamu Goto; Jacques J G H M Bergman; Naohisa Yahagi; Bas L A M Weusten
Journal:  Clin Endosc       Date:  2015-05-29

7.  Is the Endoscopic Grasp-and-Traction Device Useful for Endoscopic Submucosal Dissection in Treating Early Gastric Cancer?

Authors:  Joo Young Cho
Journal:  Clin Endosc       Date:  2015-05-29

8.  Comparable Data Between Double Endoscopic Intraluminal Operation and Conventional Endoscopic Submucosal Dissection for Esophageal Cancer.

Authors:  Makoto Sohda; Kengo Kuriyama; Tomonori Yoshida; Yuji Kumakura; Hiroaki Honjo; Makoto Sakai; Tatsuya Miyazaki; Hiroyuki Kuwano
Journal:  J Gastrointest Surg       Date:  2019-02-13       Impact factor: 3.452

9.  Efficacy and safety of an internal magnet traction device for endoscopic submucosal dissection: ex vivo study in a porcine model (with video).

Authors:  Akira Dobashi; Andrew C Storm; Louis M Wong Kee Song; Christopher J Gostout; Jodie L Deters; Charles A Miller; Mary A Knipschield; Elizabeth Rajan
Journal:  Surg Endosc       Date:  2018-10-23       Impact factor: 4.584

10.  A novel magnetic countertraction device for endoscopic submucosal dissection significantly reduces procedure time and minimizes technical difficulty.

Authors:  Hiroyuki Aihara; Marvin Ryou; Nitin Kumar; Michele B Ryan; Christopher C Thompson
Journal:  Endoscopy       Date:  2014-02-26       Impact factor: 10.093

View more