Anna Maria Di Nallo1, Lidia Strigari, Claudia Giliberti, Angelico Bedini, Raffaele Palomba, Marcello Benassi.
Abstract
BACKGROUND: The paper reports the electric, magnetic and electromagnetic fields (emf) measurements carried out in the Regina Elena National Cancer Institute (NCI). Several devices, used in diagnostics and in medical cures, can represent sources of emf for the workers and for the public subjected to the treatments. The aim is to evaluate their exposition, in order to assess the compliance with the law.Entities:
Mesh:
Year: 2008 PMID: 18598357 PMCID: PMC2478644 DOI: 10.1186/1756-9966-27-16
Source DB: PubMed Journal: J Exp Clin Cancer Res ISSN: 0392-9078
Equipments investigated in each division of the NCI-IRE
| Electro-stimulators | Biorem TENS | |
| Electro-stimulators | Biorem ionophoresis (IP) | |
| Laser therapy | Leve Laser | |
| Ultrasound therapy | Radar Biorem | |
| Cyclette | Technogym | |
| Pressotherapy | Linfopress Fisioline | |
| Lymphatic Drainage | Linfopress | |
| Lymphatic Drainage | Linfopress studio | |
| Workplaces – PC | Philips | |
| Electric panel | FAAE | |
| Monitor | Philips | |
| Ventilator | Dräger | |
| Pump | Abbott | |
| Ventilator | Dräger | |
| Electric scalpel | Martin/Conmed | |
| Magnetic Resonance | Philips (0,5 T) | |
| DECT | Alcatel |
The magnetic induction measurements near the tapis roulant (fig.1)
| 1 | |
| 2 | 0,03 |
| 3 | 0,01 |
| 4 | 0,03 |
| 5 | 0,02 |
The magnetic induction B(μT) and effective electric field Eeff(V/m) levels in the box described in fig. 2
| 1 | 0,2 | 0,2 | 0,5 | 0,1 | <1 |
| 2 | 0,1 | 0,1 | 0,5 | 0,2 | 2,3 |
| 3 | 0,07 | 0,1 | 1,1 | 0,1 | 3 |
| 4 | 0,1 | 0,1 | 0,5 | 0,3 | < 1 |
| 5 | 0,3 | 1,0 | 0,4 | 2,2 | |
| 6 | 0,02 | 0,03 | 0,3 | 0,5 | < 1 |
| 7 | 0,02 | 0,02 | 0,03 | 0,4 | < 1 |
| 8 | 0,03 | 0,02 | 0,07 | 0,3 | < 1 |
Figure 1Points of measurements near the tapis roulant (detected frequency: 50 Hz).
Figure 2Specific equipments in a physiotherapy box.
Figure 3Results of the magnetic induction measurements near the press-therapy equipments.
The B(μT) levels measured in the points reported in Fig. 3
| 1 | |
| 2 | 5,3 |
| 3 | 2,0 |
| 4 | 4,8 |
| 5 | |
| 6 | 2,3 |
| 7 | 1,1 |
The B(μT) measurements in the open space reported in Fig. 4
| H 1 m | H 1.9 m | |
| 1 | 1,7 | 1,2 |
| 2 | 1,9 | 1,1 |
| 3 | 0,9 | 0,7 |
| 4 | 0,3 | 0,2 |
| 5 | 0,3 | 0,1 |
| 6 | 0,1 | 0,1 |
| 7 | 0,07 | 0,07 |
Figure 4Results of the magnetic induction measurements B(μT) in the intensive care division.
Figure 5Measurements near a patient's bed in the intensive care division.
Figure 6Results of the spectrum analysis carried out in the physiotherapy division.
Limits, levels and quality targets for people exposure to 100 kHz-300 GHz [8]
| 0,1 < f ≤ 3 MHz | 60 | 0,2 | - |
| 3 < f ≤ 3000 MHz | 20 | 0,05 | 1 |
| 3 < f ≤ 300 GHz | 40 | 0,01 | 4 |
| 0,1 MHz< f ≤ 300 GHz | 6 | 0,016 | 0,10 (3 MHz-300 GHz) |
| 0,1 MHz< f ≤ 300 GHz | 6 | 0,016 | 0,10 (3 MHz-300 GHz) |
Limits, levels and quality targets for people exposure to power lines at 50 Hz [9]
| Eeff (kV/m) | 5 | - | - |
| Beff (μT) | 100 | 10 | 3 |
Reference levels for the general public from [11].
| 0–1 Hz | - | 3,2 × 104 | 4 × 104 | - |
| 1–8 Hz | 10 000 | 3,2 × 104/f2 | 4 × 104/f2 | - |
| 8–25 Hz | 10 000 | 4 000/f | 5000/f | - |
| 0,025–0,8 Hz | 250/f | 4/f | 5/f | - |
| 0,8–3 kHz | 250/f | 5 | 6,25 | - |
| 3–150 kHz | 87 | 5 | 6,25 | - |
| 0,15–1 MHz | 87 | 0,73/f | 0,92/f | - |
| 1–10 MHz | 87/f1/2 | 0,73/f | 0,92/f | - |
| 10–400 MHz | 28 | 0,073 | 0,092 | 2 |
| 400–2 000 MHz | 1,375 f1/2 | 0,0037 f1/2 | 0,0046 f1/2 | f/200 |
| 2–300 GHz | 61 | 0,16 | 0,20 | 10 |
The B(μT) measurements around the bed reported in Fig. 5
| 8 | 0,6 |
| 9 | 0,4 |
| 10 | 0,1 |
| 11 | |
| 12 | 1,0 |
| 13 | 0,1 |
| 14 | 6,0 |
| 15 | 0,7 |
| 16 |