| Literature DB >> 18597692 |
Chadi H Stephan1, Michel Fournier, Pauline Brousseau, Sébastien Sauvé.
Abstract
BACKGROUND: A routine method for the quantification of beryllium in biological fluids is essential for the development of a chelation therapy for Chronic Beryllium Disease (CBD). We describe a procedure for the direct determination of beryllium in undigested micro quantities of human blood and serum using graphite furnace atomic absorption spectrometry. Blood and serum samples are prepared respectively by a simple 8-fold and 5-fold dilution with a Nash Reagent. Three experimental setups are compared: using no modifier, using magnesium nitrate and using palladium/citric acid as chemical modifiers.Entities:
Year: 2008 PMID: 18597692 PMCID: PMC2459166 DOI: 10.1186/1752-153X-2-14
Source DB: PubMed Journal: Chem Cent J ISSN: 1752-153X Impact factor: 4.215
Figure 1Thermal pre-treatment pyrolysis and atomization curves for 5-fold diluted non spiked serum and 8-fold diluted non spiked blood without the addition of any chemical modifiers.
Figure 2Influence of palladium mass on the relative absorbance of beryllium in 8-fold diluted blood and 5-fold diluted serum samples.
Figure 3Variations in the background signal; (a) blood without modifier; (b) blood with 15 μg of Palladium; (c) blood with 10 μg Magnesium; (d) serum without modifier; (e) serum with 10 μg Palladium; (f) serum with 12 μg Magnesium.
Figure 4Influence of magnesium mass on the relative absorbance of beryllium in 8-fold diluted blood and 5-fold diluted serum samples.
Beryllium contents in blood and serum of ten individuals. Averages of triplicate measurements with standard deviations are given.
| Sample | Sex | Smoker | Blood (μg L-1) | Serum (μg L-1) |
| 1 | F† | SŦ | 0.71 ± 0.09 | 0.49 ± 0.08 |
| 2 | F | NSŧ | 0.64 ± 0.09 | 0.43 ± 0.09 |
| 3 | F | S | 0.74 ± 0.06 | 0.45 ± 0.08 |
| 4 | F | NS | 0.67 ± 0.08 | 0.40 ± 0.08 |
| 5 | F | NS | 0.57 ± 0.06 | 0.46 ± 0.09 |
| 6 | M | NS | 0.68 ± 0.04 | 0.40 ± 0.05 |
| 7 | F | NS | 0.64 ± 0.08 | 0.41 ± 0.04 |
| 8 | F | NS | 0.67 ± 0.09 | 0.42 ± 0.04 |
| 9 | F | NS | 0.48 ± 0.06 | 0.43 ± 0.03 |
| 10 | M‡ | NS | 0.55 ± 0.07 | 0.42 ± 0.06 |
| Mean | 0.63 ± 0.08 | 0.43 ± 0.03 |
† Female; ‡ Male; Ŧ Smoker; ŧ non smoker.
Instrument conditions and furnace programme for the determination of beryllium (argon flow rate was set at 0.3 L min-1, except during atomization)
| Step | Temp (°C) | Time (s) |
| Drying | 85 | 5 |
| Drying | 95 | 30 |
| Drying | 120 | 20 |
| Pre-pyrolysis | 450 | 22 |
| Pyrolysis | 1000†; 800‡ | 17 |
| Atomizing | 2900†; 2800‡ | 3 |
| Cleaning | 2900 | 2 |
† Serum analysis