| Literature DB >> 18585857 |
Celal Duran1, Hasan Basri Senturk, Latif Elci, Mustafa Soylak, Mehmet Tufekci.
Abstract
A new method for the preconcentration of some trace metals (Co, Ni, Cu, and Cd) as complexed with ammonium pyrrolidynedithiocarbamate (APDC) was developed using a mini-column filled with Amberlite XAD-2000 resin. Metal contents were determined by flame atomic absorption spectrometry (FAAS) after the metal complexes accumulated on the resin were eluted with 1M HNO(3) in acetone. The effects of the analytical parameters such as sample pH, quantity of complexing agent, eluent type, resin quantity, sample volume, sample flow rate, and matrix ions were investigated on the recovery of the metals from aqueous solutions. The relative standard deviation (R.S.D.) of the method was <6%. The validation of the method was confirmed using two certified reference materials (CRM TMDW-500 Drinking Water and CRM SA-C Sandy Soil C). The method was successfully applied to some stream waters and mushroom samples from Eastern Black Sea Region (Trabzon city) of Turkey.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18585857 DOI: 10.1016/j.jhazmat.2008.05.034
Source DB: PubMed Journal: J Hazard Mater ISSN: 0304-3894 Impact factor: 10.588