Literature DB >> 18582809

Continuous hemodynamic monitoring in patients with pulmonary arterial hypertension.

Robert P Frantz1, Raymond L Benza, Barbro Kjellström, Robert C Bourge, Robyn J Barst, Tom D Bennett, Michael D McGoon.   

Abstract

BACKGROUND: The purpose of this study was to determine whether an implantable hemodynamic monitor (IHM) could be used to judge the response of pulmonary arterial hypertension (PAH) patients to changes in therapy.
METHODS: A prospective, non-randomized, multi-center study evaluated physical examination, functional class, echocardiography, brain natriuretic peptide (BNP) levels, exercise capacity assessed by 6-minute walk and cardiopulmonary exercise tests, and quality of life at baseline and at 12 weeks. IHM measurements were continuously available to clinicians between clinic visits. Based on a priori, pre-specified analyses, the relationships between hemodynamic values, PAH treatments and clinical parameters were tracked in an observational fashion.
RESULTS: Twenty-four PAH patients underwent IHM implantation prior to a change in PAH therapy. IHM data identified 13 of the 15 patients who improved their 6-minute walk distance by >30 m at 12 weeks (+48 +/- 65 m, p < 0.05), whereas the others walked less (-78 +/- 115 m, not statistically significant). In addition, peak Vo(2), BNP levels and Minnesota Living with Heart Failure Questionnaire scores only improved in the former group. The change in mean pulmonary artery pressure correlated with the change in 6-minute walk distance at 12 weeks (r = -0.71, p < 0.001). Device-related adverse events were comparable to those known to occur with a pacemaker-like device.
CONCLUSIONS: Changes in ambulatory continuous hemodynamic measurements predicted changes in 6-minute walk distance after the start or addition of PAH therapy. The IHM also identified patients who had improved exercise tolerance, BNP levels and quality of life. The IHM appeared to be well tolerated and allowed rapid hemodynamic feedback between clinic visits.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18582809     DOI: 10.1016/j.healun.2008.04.009

Source DB:  PubMed          Journal:  J Heart Lung Transplant        ISSN: 1053-2498            Impact factor:   10.247


  7 in total

1.  Subject of the year: remote monitoring.

Authors:  Fred Kusumoto; Nora Goldschlager
Journal:  J Interv Card Electrophysiol       Date:  2009-04-24       Impact factor: 1.900

2.  Pulmonary hypertension: diagnosis and management.

Authors:  Michael D McGoon; Garvan C Kane
Journal:  Mayo Clin Proc       Date:  2009-02       Impact factor: 7.616

3.  A simple echocardiographic prediction rule for hemodynamics in pulmonary hypertension.

Authors:  Alexander R Opotowsky; Jason Ojeda; Frances Rogers; Vikram Prasanna; Mathieu Clair; Lilamarie Moko; Anjali Vaidya; Jonathan Afilalo; Paul R Forfia
Journal:  Circ Cardiovasc Imaging       Date:  2012-08-22       Impact factor: 7.792

Review 4.  Invasive monitoring in patients with heart failure.

Authors:  Saima Husain; Salpy V Pamboukian; Josè A Tallaj; David C McGiffin; Robert C Bourge
Journal:  Curr Cardiol Rep       Date:  2009-05       Impact factor: 2.931

Review 5.  Remote monitoring of patients with implanted cardiac devices.

Authors:  Fred Kusumoto; Nora Goldschlager
Journal:  Clin Cardiol       Date:  2010-01       Impact factor: 2.882

6.  Physical activity limitation as measured by accelerometry in pulmonary arterial hypertension.

Authors:  Meredith E Pugh; Maciej S Buchowski; Ivan M Robbins; John H Newman; Anna R Hemnes
Journal:  Chest       Date:  2012-12       Impact factor: 9.410

Review 7.  A Contemporary Systematic Approach to Assessing the Patient with Heart Failure with Reduced Ejection Fraction: Multimodal Noninvasive and Invasive Evaluation.

Authors:  Siu-Hin Wan; Paul M McKie; John P Bois
Journal:  Cardiol Res Pract       Date:  2019-09-11       Impact factor: 1.866

  7 in total