| Literature DB >> 18574630 |
T Jarzembowski1, K Wiśniewska, A Józwik, E Bryl, J Witkowski.
Abstract
We studied the usefulness of flow cytometry for detection of penicillin resistance in E. faecalis and S. aureus by direct binding of commercially available fluorescent penicillin, Bocillin FL, to cells obtained from culture. There were significantly lower percentages of fluorescent cells and median and mean fluorescence values per particle in penicillin-resistant than in penicillin-sensitive strains of both species observed. The method allows rapid detection of penicillin resistance in S. aureus and E. faecalis. The results encourage further investigations on the detection of antibiotic resistance in bacteria using flow cytometry.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18574630 DOI: 10.1007/s00284-008-9179-8
Source DB: PubMed Journal: Curr Microbiol ISSN: 0343-8651 Impact factor: 2.188