Meridional differences in internal dimensions of the anterior segment in human eyes evaluated with 2 imaging systems.

Liliana Werner1, Carlo Lovisolo, Jesse Chew, Manfred Tetz, Matthias Müller.   

Abstract

PURPOSE: To evaluate meridional differences in angle-to-angle (A-A) and sulcus-to-sulcus (S-S) dimensions in human eyes using very high frequency ultrasound (VHF US) and anterior segment optical coherence tomography (AS-OCT).
SETTING: John A. Moran Eye Center, University of Utah, USA; Berlin Eye Research Institute, Germany; Quattroelle Centri Oftalmochirurgici, Italy.
METHODS: Human autopsy eyes were fixated to special eye holders and had anterior segment scanning using 50 MHz ultrasound (Artemis, Ultralink) (20 eyes) and AS-OCT (SL-OCT, Heidelberg Engineering) (24 eyes). Seventy-three eyes (58 patients) also had anterior segment scanning using the same systems (Artemis, 50 eyes; SL-OCT, 23 eyes). Four meridians were evaluated in each eye. The A-A dimensions were measured by VHF US and AS-OCT. The S-S dimensions were also measured by VHF US. Statistical analyses were done with the Huynh-Feldt test.
RESULTS: In the autopsy-eye study, the differences in A-A and S-S dimensions between the 4 meridians in eyes analyzed with VHF US were statistically significant (P=.002 and P=.009, respectively). The differences in A-A dimensions in eyes analyzed with the AS-OCT were also significant (P=.005). The same findings were observed in the clinical study (VHF US, P=.001 for A-A and P<.001 for S-S; AS-OCT, P<.001 for A-A).
CONCLUSIONS: This study suggests that the internal geometry of the anterior segment of human eyes does not constitute a circle. This may have significant relevance for the accurate sizing of phakic intraocular lenses.

Entities:  

Mesh:

Year:  2008        PMID: 18571080     DOI: 10.1016/j.jcrs.2008.03.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

2.  Causes of Phakic Implantable Collamer Lens Explantation/Exchange at King Khaled Eye Specialist Hospital.

Authors:  Nasser A AlSabaani; Ashley Behrens; Sabah Jastanieah; Salem Al Malki; Mohanna Al Jindan; Saeed Al Motowa
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Oct-Dec

3.  Changes in symmetry of anterior chamber following routine cataract surgery in non-glaucomatous eyes.

Authors:  Hanbin Lee; Ieva Zukaite; Valerie Juniat; Maria E Dimitry; Amanda Lewis; Mayank A Nanavaty
Journal:  Eye Vis (Lond)       Date:  2019-07-03