Literature DB >> 18567301

Development and validation of a stability-indicating column high-performance liquid chromatographic assay method for determination of nebivolol in tablet formulation.

Pankaj K Kachhadia1, Ashish S Doshi, Hitendra S Joshi.   

Abstract

A simple, precise, and accurate isocratic reversed-phase (RP) stability-indicating column high-performance liquid chromatographic (HPLC) assay method was developed and validated for determination of nebivolol in solid pharmaceutical dosage forms. Isocratic RP-HPLC separation was achieved on a Phenomenex Luna C8 (2) column (250 mm x 4.6 mm id, 5 microm particle size) using mobile phase composed of acetonitrile-pH 3.5 phosphate buffer (35 + 65, v/v) at a flow rate of 1.0 mL/min, and detection was performed at 280 nm using a photodiode array detector. The drug was subjected to oxidation, hydrolysis, photolysis, and heat to apply stress conditions. The method was validated for specificity, linearity, precision, accuracy, robustness, and solution stability. The method was linear in the drug concentration range of 40-160 microg/mL with a correlation coefficient of 0.9999. The repeatability relative standard deviation (RSD) for 6 samples was 0.69%, and the intermediate precision (RSD) for 6 samples was 1.39%. The accuracy (recovery) was between 98.57 and 99.55%. Degradation products produced as a result of stress studies did not interfere with detection of nebivolol, and the assay can thus be considered stability-indicating.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18567301

Source DB:  PubMed          Journal:  J AOAC Int        ISSN: 1060-3271            Impact factor:   1.913


  1 in total

1.  Simultaneous estimation of amlodipine besylate and nebivolol hydrochloride in tablet dosage forms by reverse phase-high-performance liquid chromatographic using ultraviolet detection.

Authors:  Deepak Sharma; Anurekha Jain; Alankar Shrivastava
Journal:  Pharm Methods       Date:  2011-01