Literature DB >> 18564214

A new vascular closure device for the transradial approach: the D Stat Radial system.

Hubertus von Korn1, Marc A Ohlow, Jiangtao Yu, Burkhard Huegl, Walter Schulte, Andreas Wagner, Kai Haberl, Stefan Gruene, Bernward Lauer.   

Abstract

BACKGROUND: The transradial approach is associated with low complication rates. The D Stat Radial vascular closure system offers hemostatic pressure locally at the puncture site with residual venous flow.
METHODS: We prospectively included 113 consecutive patients presenting between August 2006 and December 2006. Diagnostic coronary procedures were performed using 4 Fr or 5 Fr sheaths, while 6 Fr devices were used for PCI. In every case at least 5.000 IU heparin was given. Compression with the retention strap was planned for approximately 3 hours before the pad was fixed using a medical strip. All data were entered into a database after a two-dimensional ultrasound examination study of the puncture site.
RESULTS: The mean age was 65 years (+/-SD 10.3), with 62% being male. Patients presenting with acute coronary syndrome amounted to 6.2%, where 5r F sheaths were usually used with these patients (64.6%). PCI was carried out in 21.2% of cases. The system was applied successfully in all patients. The mean duration of compression was 4.6 hours (+/-SD 1.4). Bleeding after removal of the system occurred in 18.6% of cases, which resulted in prolonged compression in 17.7%. A hematoma >5 cm was seen in 4.4% of the sample. Ultrasound examinations revealed closure of the radial artery in 6.2%.
CONCLUSIONS: We conclude that this new device operates efficiently, but bleeding occurred in 18.6%. Compression time should be extended to 6 hours.

Entities:  

Mesh:

Year:  2008        PMID: 18564214     DOI: 10.1111/j.1540-8183.2008.00366.x

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  2 in total

1.  [Prevention of pocket-related complications following heart rhythm device implantation. D-Stat Hemostat™ versus vacuum drainage].

Authors:  Marc-Alexander Ohlow; Björn Buchter; Michele Brunelli; Bernward Lauer; Matthias Schreiber; J Christoph Geller
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-02-03

2.  Revealing the impact of local access-site complications and upper extremity dysfunction post transradial percutaneous coronary procedures.

Authors:  E M Zwaan; A G M M Koopman; C A J Holtzer; F Zijlstra; M J P F Ritt; G Amoroso; E Moerman; M J M Kofflard; A A J IJsselmuiden
Journal:  Neth Heart J       Date:  2015-11       Impact factor: 2.380