| Literature DB >> 18561939 |
Kouwa Yamashita1, Mitsunobu Okuyama, Risa Nakagawa, Seijiro Honma, Fumitoshi Satoh, Ryo Morimoto, Sadayoshi Ito, Madoka Takahashi, Mitsuteru Numazawa.
Abstract
A highly sensitive quantification method of aldosterone by liquid chromatography-electrospray ionization tandem mass spectrometry (LC-ESI-MS/MS) was investigated in a positive mode using recently developed picolinyl derivatization. Aldosterone was smoothly and quantitatively converted to the ethyl ether-picolinyl derivative by treatment with HCl-ethanol followed by the esterification with picolinic acid in the presence of 2-methyl-6-nitrobenzoic anhydride and 4-dimethylaminopyridine. The positive ion-ESI mass spectrum of the ethyl ether-picolinyl derivative was characterized by an appearance of protonated molecule ([M+H](+)) as a base peak. The ethyl ether-picolinyl derivatization gave a successful result in a separation of aldosterone from corticosterone, dehydrocorticosterone and cortexolone, and also provided an approximately 10-fold higher ESI response in the positive-LC-ESI-MS/MS (selected reaction monitoring; SRM) when compared to that of underivatized molecule (negative mode). The limit of quantification of aldosterone by SRM using ethyl ether-picolinyl derivatization (m/z 494-->m/z 448) was 1 pg/0.2 ml serum with accuracy and precision of 92.6% and 5.6%, respectively.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18561939 DOI: 10.1016/j.chroma.2008.05.034
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759