Teng-Yeow Tan1, Yun-Yu Lin, Ulf Schminke, Ting-Yao Chen.
Abstract
Carotid duplex sonography (CDS) is regarded as a screening tool for lateral dural arteriovenous fistulae (AVF). However, data on evaluating long-term effects of endovascular treatment are limited. We report the CDS findings in the feeding arteries of a traumatic temporal extradural AVF before and after transarterial embolization. Volume flow of the left common carotid artery was greater than the right (433 ml/minute versus 294 ml/minute right) and the resistance index of the left external carotid artery was lower than the right (0.69 left versus 0.84 right). Both parameters returned to normal 4 months after embolization, thus confirming successful endovascular treatment. (c) 2008 Wiley Periodicals, Inc.Entities:
Mesh:
Year: 2008 PMID: 18561342 DOI: 10.1002/jcu.20505
Source DB: PubMed Journal: J Clin Ultrasound ISSN: 0091-2751 Impact factor: 0.910