Literature DB >> 18560937

Late recurrence of a hepatic artery aneurysm after treatment using an endovascular stent.

Jonathan Downer1, Kiyoshi Choji.   

Abstract

Endovascular stent placement and coil embolization have become established options in the treatment of visceral arterial aneurysms. In this article we report the case of an 83-year-old presenting with gastrointestinal hemorrhage due to a recurrent hepatic arterial aneurysm occurring 12 years after treatment with an endovascular stent. The recurrent aneurysm had resulted from stent fracture and was successfully treated by coil embolization. To our knowledge, stent fracture complicating the endovascular treatment of a visceral artery aneurysm has not been described in the published literature. With the increasing use of metallic endoprostheses in interventional radiology, recognizing and reporting device failure are of critical importance.

Entities:  

Mesh:

Year:  2008        PMID: 18560937     DOI: 10.1007/s00270-008-9368-7

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  1 in total

1.  Multiphase multidetector CT in the diagnosis of haemobilia: a potentially catastrophic ruptured hepatic artery aneurysm complicating the treatment of a patient with locally advanced rectal cancer.

Authors:  A M Mortimer; A Wallis; A Planner
Journal:  Br J Radiol       Date:  2011-05       Impact factor: 3.039

  1 in total