Literature DB >> 1855801

A beamformer for the acquisition of evoked potentials.

C A McKinley1, P A Parker.   

Abstract

Evoked potentials (EP) contain information about various physiological parameters and the estimation and detection of these signals can aid in the diagnosis of many pathological conditions. However, the signal-to-noise ratio (SNR) for EP measurement is often very low, and thus signal processing techniques must be employed to enhance the SNR. A delay and sum beamformer acquisition system has the potential for significant SNR improvement in EP measurements. In this communication it is shown that an electrode array acquisition system implements a uniform coherent delay and sum beamformer. The performance of the beamformer is characterized in terms of the number of electrodes, and cross-channel correlation. When compared to conventional ensemble averaging, the beamformer reduces the number of response repetitions required to achieve a given SNR by a factor which approaches the number of channels in the acquisition system.

Mesh:

Year:  1991        PMID: 1855801     DOI: 10.1109/10.133234

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

Review 1.  Methods for estimating muscle fibre conduction velocity from surface electromyographic signals.

Authors:  D Farina; R Merletti
Journal:  Med Biol Eng Comput       Date:  2004-07       Impact factor: 2.602

2.  Noise characteristics of stainless-steel surface electrodes.

Authors:  D T Godin; P A Parker; R N Scott
Journal:  Med Biol Eng Comput       Date:  1991-11       Impact factor: 2.602

  2 in total