Light scatter on the surface of AcrySof intraocular lenses: part I. Analysis of lenses retrieved from pseudophakic postmortem human eyes.

Shigeo Yaguchi1, Hitoshi Nishihara, Waraporn Kambhiranond, Daniel Stanley, David J Apple.   

Abstract

BACKGROUND AND
OBJECTIVE: To investigate the cause of light scatter measured on the surface of AcrySof intraocular lenses (Alcon Laboratories, Inc., Fort Worth, TX) retrieved from pseudophakic postmortem human eyes.
MATERIALS AND METHODS: Ten intraocular lenses (Alcon AcrySofModel MA60BM) were retrieved postmortem and analyzed for light scatter before and after removal of surface-bound biofilms.
RESULTS: Six of the 10 lenses exhibited light scatter that was clearly above baseline levels. In these 6 lenses, both peak and average pixel density were reduced by approximately 80% after surface cleaning.
CONCLUSION: The current study demonstrates that a coating deposited in vivo on the lens surface is responsible for the light scatter observed when incident light is applied.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18556941     DOI: 10.3928/15428877-20080501-22

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  4 in total

1.  Effects of glistenings in intraocular lenses.

Authors:  Marrie van der Mooren; Luuk Franssen; Patricia Piers
Journal:  Biomed Opt Express       Date:  2013-07-11       Impact factor: 3.732

2.  Decreased visual function due to high-level light scattering in a hydrophobic acrylic intraocular lens.

Authors:  Shinichiro Yoshida; Hiroyuki Matsushima; Mayumi Nagata; Tadashi Senoo; Ichiro Ota; Kensaku Miyake
Journal:  Jpn J Ophthalmol       Date:  2011-02-18       Impact factor: 2.447

3.  Influence of surface light scattering in hydrophobic acrylic intraocular lenses on laser beam transmittance.

Authors:  Tomoyasu Shiraya; Satoshi Kato; Keiichiro Minami; Kazunori Miyata
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

4.  Comparison of glistenings formation and their effect on forward light scatter between the Acrysof SN60WF and Eternity Natural Uni NW-60 intraocular lenses.

Authors:  Eva Philippaki; David P O'Brart; Chistopher C Hull
Journal:  BMJ Open Ophthalmol       Date:  2020-02-12