| Literature DB >> 18555651 |
Masashi Tamaki1, Tetsuya Tomita, Takaharu Yamazaki, William J Hozack, Hideki Yoshikawa, Kazuomi Sugamoto.
Abstract
The objective of this study was to evaluate in vivo kinematics of a high-flexion, posterior-stabilized fixed-bearing, total knee arthroplasty in weight-bearing deep knee-bending motion. A total of 20 knees implanted with the Scorpio Non-Restrictive Geometry knee system in 17 patients were assessed in this study. The Scorpio Non-Restrictive Geometry is a recent implant design with modifications made to accommodate a higher flexion range of motion and greater axial rotation, particularly during more functionally demanding activities. Patients were examined during a deep knee-bending motion using fluoroscopy, and femorotibial motion was determined using a 2-dimensional to 3-dimensional registration technique. The average flexion angle was 126.5 degrees (110 degrees -149 degrees ). The femoral component demonstrated a mean of 13.5 degrees (5.2 degrees -21 degrees ) external rotation. The external rotation increased up to maximum flexion. The pivot pattern was a medial pivot pattern similar to that reported in normal knee kinematics.Entities:
Mesh:
Year: 2008 PMID: 18555651 DOI: 10.1016/j.arth.2008.04.009
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757