Microbial validation of vent filters.

J H Robertson1, W R Frieben.   

Abstract

The Upjohn Company uses filtration to remove microorganisms and particulates from air and other gases which may come in contact with sterile products. To validate the microbial retentivity of these filters, they were challenged with an aerosol of Bacillus subtilis var niger spores. An aerosol challenge was used because it more closely simulated the use for which these filters were designed. The test apparatus was constructed of autoclavable components using a jet-type nebulizer and heated air mixing tube. Characterization of the aerosol particle size distribution with a particle size analyzer demonstrated that 80% of the particles had a diameter of x 3.0 times;m and that the particles had a mean mass diameter of 1.9 times;m with a geometric standard deviation of 1.8 times;m. Studies conducted with aerosols of Bacillus subtilis var niger spores demonstrated that the test apparatus could recover ca. 50% of the spores that were aerosolized. Hydrophobic filters from various manufactures were challenged with an aerosol of at least 10(8) spores of Bacillus subtilis. All filters tested could retain at least 10(9) spores when physical integrity of the filter was verfield.

Entities:  

Year:  1984        PMID: 18553465     DOI: 10.1002/bit.260260803

Source DB:  PubMed          Journal:  Biotechnol Bioeng        ISSN: 0006-3592            Impact factor:   4.530


  1 in total

1.  Monitoring of aerial pollutants emitted from Swine houses in Korea.

Authors:  Ki Y Kim; Han J Ko; Hyeon T Kim; Yoon S Kim; Young M Roh; Cheol M Lee; Chi N Kim
Journal:  Environ Monit Assess       Date:  2007-02-01       Impact factor: 2.513

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.