Literature DB >> 18550476

Accuracy of capillary whole blood international normalized ratio on the CoaguChek S, CoaguChek XS, and i-STAT 1 point-of-care analyzers.

Brad S Karon1, Robert D McBane, Rajeev Chaudhry, Lisa K Beyer, Paula J Santrach.   

Abstract

We evaluated the accuracy of capillary whole blood international normalized ratio (INR) on the CoaguChek S (Roche Diagnostics, Indianapolis, IN), CoaguChek XS (Roche Diagnostics), and i-STAT 1 (i-STAT, East Windsor, NJ) point-of-care (POC) analyzers compared with venous plasma INRs determined by a reference laboratory method. Overall agreement between POC and laboratory plasma INR was very good, with median bias between capillary whole blood and laboratory plasma INRs varying from 0.0 to -0.2 INR units on all devices. More than 90% of results on the CoaguChek XS and i-STAT 1 and 88% of CoaguChek S results were within 0.4 INR units of the reference laboratory method. The CoaguChek XS and i-STAT 1 demonstrated greater accuracy than the CoaguChek S as measured by the number of results that differed by more than 0.5 INR units from the reference method. Median bias between CoaguChek S capillary whole blood and laboratory plasma INRs changed over time, demonstrating the need for ongoing quality assurance measures for POC INR programs.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18550476     DOI: 10.1309/DEK41W141Y0KRN8A

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  5 in total

Review 1.  [Portable coagulometer devices in the monitoring and control of oral anticoagulation therapy: a systematic review].

Authors:  Javier Caballero-Villarraso; Román Villegas-Portero; Fernando Rodríguez-Cantalejo
Journal:  Aten Primaria       Date:  2010-10-30       Impact factor: 1.137

2.  Administration of tissue plasminogen activator without coagulation results in a Chinese population.

Authors:  Xiaoming Qin; Songyao Zhao; Liujie Yin; Hailing Dou; Jie Chen; Yifan Wang; Mingzhe Li; Ruifang Chen; Jing Fu; Wei Liu; Xin Liu; Gaiqing Yang; Runqing Wang; Xinzhou Jia; Shufang Bu; Dongpu Ma; Baoyu Wang; Shize Li
Journal:  Neurol Sci       Date:  2018-01-03       Impact factor: 3.307

3.  All the bang without the bucks: Defining essential point-of-care testing for traumatic coagulopathy.

Authors:  Michael D Goodman; Amy T Makley; Dennis J Hanseman; Timothy A Pritts; Bryce R H Robinson
Journal:  J Trauma Acute Care Surg       Date:  2015-07       Impact factor: 3.313

4.  Trends in Warfarin Monitoring Practices Among New York Medicare Beneficiaries, 2006-2011.

Authors:  Darren M Triller; Susan Wymer; Patrick D Meek; Elaine M Hylek; Jack E Ansell
Journal:  J Community Health       Date:  2015-10

5.  In-home Compared With In-Clinic Warfarin Therapy Monitoring in Mechanical Heart Valves: A Population-Based Study.

Authors:  Martin van Zyl; Waldemar E Wysokinski; Thomas M Jaeger; Ana I Casanegra; Bernard J Gersh; Robert D McBane
Journal:  Mayo Clin Proc Innov Qual Outcomes       Date:  2020-08-15