Contribution of morphology discrimination algorithm for improving rhythm discrimination in slow and fast ventricular tachycardia zones in dual-chamber implantable cardioverter-defibrillators.

Giuseppe Boriani1, Eraldo Occhetta, Silvio Cesario, Stefano Grossi, Marco Marconi, Giancarlo Speca, Paolo Silvestri, Mauro Biffi, Miriam Bortnik, Cristian Martignani, Angelo Branzi.   

Abstract

AIMS: Morphology discrimination (MD) is an algorithm based on QRS morphology analysis, that can be used alone or in various combinations with other discriminators in order to diagnose ventricular tachycardia (VT) [the implantable cardioverter-defibrillator (ICD) may be set to diagnose VT if 'Any' or 'All' the discriminators indicate VT)]. METHODS AND
RESULTS: We evaluated the contribution of MD in terms of Specificity (SP) and Sensitivity (SE) of rhythm discrimination in slow and fast VT zones (rates <150 bpm and between 150 and 180 bpm, respectively). Detection results (682 spontaneous episodes in 58 patients) were corrected for multiple episodes within a patient (generalized estimating equations method). Rhythm discrimination in slow VT zones was characterized by SE and SP below 80% without MD, while the use of MD alone allowed to enhance SP, maintaining SE at 96%. Use of MD in combination with other discriminators resulted in a decrease in SP (setting 'Any'). In fast VT zones, MD allows an increase in SE for VT detection from around 82% up to approximately 94%, coupled with an SP of 95-96%., although SP is much higher with MD alone (94.8%) than with any other combination (78.4%). Both in slow and fast VT zones, use of multiple discriminators with the setting 'All' resulted in an unacceptable decrease in SE.
CONCLUSION: MD in a dual-chamber ICD in combination with a rate branch classification system makes it possible to achieve a very high SP in discriminating detected events both in slow and fast VT zones, while maintaining SE for VT detection.

Entities:  

Mesh:

Year:  2008        PMID: 18541620     DOI: 10.1093/europace/eun146

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

Review 1.  ICD programming.

Authors:  Mauro Biffi
Journal:  Indian Heart J       Date:  2013-12-20

2.  Implantable cardioverter defibrillator and inappropriate therapy: "black box" examination yielded both human and technical causes.

Authors:  Antoine Kossaify
Journal:  Clin Med Insights Case Rep       Date:  2013-12-02
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.