Urinary bladder and oesophageal temperatures correlate better in patients with high rather than low urinary flow rates during non-cardiac surgery.

H Sato1, M Yamakage, K Okuyama, Y Imai, H Iwashita, T Ishiyama, T Matsukawa.   

Abstract

BACKGROUND AND
OBJECTIVE: To investigate the effect of urinary flow rate on the urinary bladder temperature, we compared the accuracy and precision of urinary bladder temperature with oesophageal temperature at both high and low urine flow rates.
METHODS: Twenty-four patients ASA physical status I or II who were undergoing tympanoplasty were randomly assigned to two groups with different intravenous fluid volumes: high (10 mL kg(-1) h(-1), n = 12) and low (3 mL kg(-1) h(-1), n = 12). General anaesthesia was induced with propofol and maintained with sevoflurane (1.5-2.5%) in nitrous oxide and oxygen. Urinary bladder temperature was measured using a Foley urinary catheter; distal oesophageal temperature was measured using a stethoscope thermocouple. These temperatures were measured every 5 min during surgery and the accuracy and precision of urinary bladder temperature with oesophageal temperature were determined using regression and Bland and Altman analyses.
RESULTS: The correlation coefficient for oesophageal and urinary bladder temperature was 0.90 in the high urinary volume group and 0.75 in the low urinary volume group. The offset (oesophageal-urinary bladder) was -0.13 +/- 0.32 degrees C and -0.46 +/- 0.45 degrees C, respectively.
CONCLUSION: Urinary bladder temperature appears to be more accurate at high urinary flow rates than at low urinary flow rates for clinical use.

Entities:  

Mesh:

Year:  2008        PMID: 18538052     DOI: 10.1017/S0265021508004602

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  2 in total

1.  Clinical evaluation of a cutaneous zero-heat-flux thermometer during cardiac surgery.

Authors:  Christophe Verheyden; Arne Neyrinck; Annouschka Laenen; Steffen Rex; Elke Van Gerven
Journal:  J Clin Monit Comput       Date:  2021-09-24       Impact factor: 1.977

2.  Accuracy of zero-heat-flux thermometry and bladder temperature measurement in critically ill patients.

Authors:  Anselm Bräuer; Albulena Fazliu; Thorsten Perl; Daniel Heise; Konrad Meissner; Ivo Florian Brandes
Journal:  Sci Rep       Date:  2020-12-10       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.