Surface damage after multiple dislocations of a 38-mm-diameter, metal-on-metal hip prosthesis.

John G Bowsher1, Thomas K Donaldson, Paul A Williams, Ian C Clarke.   

Abstract

We present, for the first time, a detailed damage assessment of a large-diameter metal-on-metal (L-MOM) hip prosthesis to show the extent of surface damage that can occur in a patient after multiple dislocations. The patient was a man (51 years old) who dislocated 8 times and was finally revised at 27 months. Radiographically, the cup was malpositioned with 65 degrees lateral opening and 15 degrees retroversion. The retrieved cup was a 1-piece, 38-mm Co-Cr-Mo (M2a; Biomet, Warsaw, Ind) with a titanium-alloy backing. The retrieved components demonstrated all known modes of wear, including a polished wear scar, multidirectional scratching, "stripe" wear, surface contamination of titanium-alloy, front face wear, and backside wear. The clinical significance is that cup positioning remains critical regardless of whether a large diameter head is used or not.

Entities:  

Mesh:

Year:  2008        PMID: 18534503     DOI: 10.1016/j.arth.2007.09.007

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  THA Retrievals: The Need to Mark the Anatomic Orientation of the Femoral Head.

Authors:  Anneliese D Heiner; Karen M Kruger; Nishant M Tikekar; John J Callaghan; John J Lannutti; Thomas D Brown
Journal:  J Arthroplasty       Date:  2015-01-23       Impact factor: 4.757

2.  In vitro evaluation of stiffness graded artificial hip joint femur head in terms of joint stresses distributions and dimensions: finite element study.

Authors:  H Fouad
Journal:  J Mater Sci Mater Med       Date:  2011-04-20       Impact factor: 3.896

3.  Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings.

Authors:  Ian C Clarke; Jean-Yves Lazennec; Adrien Brusson; Christina Savisaar; John G Bowsher; Michelle Burgett; Thomas K Donaldson
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

4.  A novel formulation for scratch-based wear modelling in total hip arthroplasty.

Authors:  Karen M Kruger; Nishant M Tikekar; Anneliese D Heiner; Thomas E Baer; John J Lannutti; John J Callaghan; Thomas D Brown
Journal:  Comput Methods Biomech Biomed Engin       Date:  2013-01-10       Impact factor: 1.763

5.  A simulator study of adverse wear with metal and cement debris contamination in metal-on-metal hip bearings.

Authors:  T Halim; I C Clarke; M D Burgett-Moreno; T K Donaldson; C Savisaar; J G Bowsher
Journal:  Bone Joint Res       Date:  2014-03       Impact factor: 5.853

6.  Encoding scratch and scrape features for wear modeling of total joint replacements.

Authors:  Karen M Kruger; Nishant M Tikekar; Anneliese D Heiner; Thomas E Baer; John J Lannutti; John J Callaghan; Thomas D Brown
Journal:  Comput Math Methods Med       Date:  2013-04-11       Impact factor: 2.238

7.  In vivo evaluation of edge-loading in metal-on-metal hip resurfacing patients with pseudotumours.

Authors:  Y-M Kwon; S J Mellon; P Monk; D W Murray; H S Gill
Journal:  Bone Joint Res       Date:  2012-04-01       Impact factor: 5.853

8.  Bearing-Foreign Material Deposition on Retrieved Co-Cr Femoral Heads: Composition and Morphology.

Authors:  Nishant M Tikekar; Anneliese D Heiner; Thomas E Baer; Karen M Kruger; John J Callaghan; Thomas D Brown; John J Lannutti
Journal:  Biomed Res Int       Date:  2015-07-05       Impact factor: 3.411

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.