Evaluation of three methods for measurement of hemoglobin and calculated hemoglobin parameters with the ADVIA 2120 and ADVIA 120 in dogs, cats, and horses.

Natali Bauer1, Andreas Moritz.   

Abstract

BACKGROUND: Besides flow cytometric detection of cellular hemoglobin (HGB) concentration, the ADVIA 2120 uses a novel cyanide-free colorimetric method to determine extracellular total HGB concentration. In human samples, the results are equivalent to those of the cyanmethemoglobin method on the ADVIA 120. Cyanide-free HGB measurement has not been evaluated in animal samples.
OBJECTIVES: The aim of this prospective study was to compare the 3 methods of HGB analysis on the ADVIA 2120 and ADVIA 120 in blood samples from dogs, cats, and horses.
METHODS: Consecutive fresh K(3)EDTA blood samples from 119 dogs, 113 cats, and 151 horses submitted to the Central Laboratory, Department of Veterinary Clinical Sciences, Justus-Liebig University Giessen, were included. A CBC was performed on each sample using the ADVIA 2120 and ADVIA 120. Colorimetric and cellular HGB concentrations and all calculated variables based on HGB measurement were compared using linear regression, Passing Bablok regression, and Bland Altman plots, using the ADVIA 120 as the reference method.
RESULTS: In samples from all species, an excellent correlation was found for colorimetric HGB results (r=0.99). HGB measured with the cyanide-free method was overestimated on the ADVIA 2120 compared with the cyanide-based method on the ADVIA 120, with a mean proportional bias of -21.0% (dog), -22.0% (cat), and -19.4% (horse). The correlation of cellular HGB concentration between analyzers was excellent (r=0.99); however, imprecision was higher than for colorimetric methods. Excellent to fair agreement was found for all calculated variables.
CONCLUSION: The cyanide-free method of HGB determination is appropriate for use in blood samples from animals, provided the proportional bias is considered.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18533916     DOI: 10.1111/j.1939-165X.2008.00039.x

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  4 in total

1.  Heinz body-related interference with leukocyte and erythrocyte variables obtained by an automated hematology analyzer in cats.

Authors:  Francesco Dondi; Kateryna Vasylyeva; Federica Serafini; Marta Gruarin; Roberta Troìa; Massimo Giunti; Chiara Agnoli
Journal:  J Vet Diagn Invest       Date:  2019-07-10       Impact factor: 1.279

2.  Evaluation of red blood cell distribution width in dogs with various illnesses.

Authors:  Carlos Martinez; Carmel T Mooney; Robert E Shiel; Pak Kan Tang; Louise Mooney; Emma J O'Neill
Journal:  Can Vet J       Date:  2019-09       Impact factor: 1.008

3.  Diagnostic value of reticulocyte indices for the assessment of the iron status of cats with chronic kidney disease.

Authors:  Adeline Betting; Ariane Schweighauser; Thierry Francey
Journal:  J Vet Intern Med       Date:  2022-01-28       Impact factor: 3.333

4.  Evaluation of the scil vCell 5, a novel laser- and impedance-based point-of-care hematology analyzer, for use in dogs and cats.

Authors:  Kim-Lina Charlotte Zelmer; Natali Bauer; Andreas Moritz
Journal:  J Vet Diagn Invest       Date:  2022-03-24       Impact factor: 1.279

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.