Literature DB >> 18533901

Innovative integrated system for real-time measurement of hybridization and melting on standard format microarrays.

Yann Marcy1, Pierre-Yves Cousin, Maxime Rattier, Gordana Cerovic, Guilhem Escalier, Gilles Béna, Maurice Guéron, Lorcan McDonagh, Françoise le Boulaire, Henri Bénisty, Claude Weisbuch, Jean-Christophe Avarre.   

Abstract

Despite the great popularity and potential of microarrays, their use for research and clinical applications is still hampered by lengthy and costly design and optimization processes, mainly because the technology relies on the end point measurement of hybridization. Thus, the ability to monitor many hybridization events on a standard microarray slide in real time would greatly expand the use and benefit of this technology, as it would give access to better prediction of probe performance and improved optimization of hybridization parameters. Although real-time hybridization and thermal denaturation measurements have been reported, a complete walk-away system compatible with the standard format of microarrays is still unavailable. To address this issue, we have designed a biochip tool that combines a hybridization station with active mixing capability and temperature control together with a fluorescence reader in a single compact benchtop instrument. This integrated live hybridization machine (LHM) allows measuring in real time the hybridization of target DNA to thousands of probes simultaneously and provides excellent levels of detection and superior sequence discrimination. Here we show on an environmental single nucleotide polymorphism (SNP) model system that the LHM enables a variety of experiments unachievable with conventional biochip tools.

Mesh:

Year:  2008        PMID: 18533901     DOI: 10.2144/000112758

Source DB:  PubMed          Journal:  Biotechniques        ISSN: 0736-6205            Impact factor:   1.993


  5 in total

1.  Salt concentration effects on equilibrium melting curves from DNA microarrays.

Authors:  J Fuchs; J-B Fiche; A Buhot; R Calemczuk; T Livache
Journal:  Biophys J       Date:  2010-09-22       Impact factor: 4.033

Review 2.  Real-time DNA microarrays: reality check.

Authors:  Alexander Chagovetz; Steve Blair
Journal:  Biochem Soc Trans       Date:  2009-04       Impact factor: 5.407

3.  Magnetoresistive sensors for measurements of DNA hybridization kinetics - effect of TINA modifications.

Authors:  G Rizzi; M Dufva; M F Hansen
Journal:  Sci Rep       Date:  2017-02-07       Impact factor: 4.379

4.  Microfludic device for creating ionic strength gradients over DNA microarrays for efficient DNA melting studies and assay development.

Authors:  Jesper Petersen; Lena Poulsen; Henrik Birgens; Martin Dufva
Journal:  PLoS One       Date:  2009-03-11       Impact factor: 3.240

5.  Direct calibration of PICKY-designed microarrays.

Authors:  Hui-Hsien Chou; Arunee Trisiriroj; Sunyoung Park; Yue-Ie C Hsing; Pamela C Ronald; Patrick S Schnable
Journal:  BMC Bioinformatics       Date:  2009-10-23       Impact factor: 3.169

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.