| Literature DB >> 18533249 |
Lingyan Zhang1, Yan Liu, Tao Chen.
Abstract
A novel experimental methodology for studying a mediatorless and label-free immunosensor is proposed by immobilizing antibody on gold nanoparticle/L-cysteine coated electrode (nano-Au/L-cysteine electrode). Differential pulse voltammograms (DPV) resulting from the assembled immunosensor indicate that the immunosensor shows excellent electrochemical response to dopamine so that the electrochemical response is utilized for the signal generation step of the immunosensor. Therefore, by means of unenzymatic-labeling procedure combined with the amperometric detection using dopamine as substrate, the immunological reaction can be detected. After the immunosensor is incubated with h-IgG solution, the access of electrocatalytic behavior center of the immunosensor to dopamine is partly inhibited, which leads to a linear decrease in amperometric response of the immunosensor with h-IgG concentration over a range 0.82-90 ng mL(-1) by DPV.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18533249 DOI: 10.1016/j.ijbiomac.2008.04.010
Source DB: PubMed Journal: Int J Biol Macromol ISSN: 0141-8130 Impact factor: 6.953