Literature DB >> 18523320

Self-expanding platinum-coated nitinol devices for transcatheter closure of atrial septal defect: prevention of nickel release.

Pornthep Lertsapcharoen1, Apichai Khongphatthanayothin, Suphot Srimahachota, Ruenreong Leelanukrom.   

Abstract

BACKGROUND: A variety of nitinol-containing devices for transcatheter closure of atrial septal defects (ASD) has been widely used. However, there is concern about the release of nickel after nitinol device implantation. In this study, a platinum-coated nitinol device was braided from nanoplatinum-coated nitinol wires in order to prevent nickel release. The serum nickel levels before and after device implantation and the 1-year results were evaluated.
METHODS: Thirty-one patients, aged 4-59 years, and weighing 13.7-90.0 kg, underwent transcatheter closure. Blood samples for serum nickel levels were taken before, 1 day, 1 week, 1 month and 3 months after implantation.
RESULTS: Twenty-nine (93.6%) patients had a successful implantation. The mean ASD diameter was 19.7 +/- 4.8 mm (range 10-30 mm). Procedure-related complications included transient brachial plexus injury in 1 patient and transient dysrhythmia in 4 patients. All 29 patients had complete closure within 1 month after implantation. The mean serum nickel levels at baseline and at 1 day, 1 week, 1 month and 3 months after implantation were 0.65 +/- 0.28, 0.63 +/- 0.18, 0.67 +/- 0.34, 0.55 +/- 0.16, 0.52 +/- 0.14 ng/ml, respectively. There was no significant difference in serum nickel levels before and after implantation. There were no device-related complications at 1-year follow up.
CONCLUSIONS: Transcatheter ASD closure using a platinum-coated nitinol device can be performed safely and successfully with good outcomes. Nano-coating of platinum on nitinol wires can prevent nickel release following device implantation.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18523320

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  3 in total

1.  Endocarditis and Incomplete Endothelialization 12 Years after Amplatzer Septal Occluder Deployment.

Authors:  Allan K Nguyen; Brian A Palafox; Joanne P Starr; Richard N Gates; Farbouch Berdjis
Journal:  Tex Heart Inst J       Date:  2016-06-01

Review 2.  Transcatheter closure of atrial septal defect: principles and available devices.

Authors:  Se Yong Jung; Jae Young Choi
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

3.  Study of Nickel Levels in Patients with Atrial Septal Defect Undergoing Amplatzer Device Closure.

Authors:  Ganesh Narayana; Ganavi Ramagopal; Bhanu Duggal; Narender Omprakash Bansal
Journal:  Heart Views       Date:  2018 Jul-Sep
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.