Computed tomographic imaging of stapes implants.

Frank M Warren1, Sterling Riggs, Richard H Wiggins.   

Abstract

HYPOTHESIS: Computed tomographic (CT) imaging of stapes prostheses is inaccurate.
BACKGROUND: Clinical situations arise in which it would be helpful to determine the depth of penetration of a stapes prosthesis into the vestibule. The accuracy of CT imaging for this purpose has not been defined. This study was aimed to determine the accuracy of CT imaging to predict the depth of intrusion of stapes prostheses into the vestibule.
METHODS: The measurement of stapes prostheses by CT scan was compared with physical measurements in 8 cadaveric temporal bones.
RESULTS: The depth of intrusion into the vestibule of the piston was underestimated in specimens with the fluoroplastic piston by a mean of 0.5 mm when compared with the measurements obtained in the temporal bones. The depth of penetration of the stainless steel implant was overestimated by 0.5 mm when compared with that in the temporal bone.
CONCLUSION: The type of implant must be taken into consideration when estimating the depth of penetration into the vestibule using CT scanning because the imaging characteristics of the implanted materials differ. The position of fluoroplastic pistons cannot be accurately measured in the vestibule. Metallic implants are well visualized, and measurements exceeding 2.2 mm increase the suspicion of otolithic impingement. Special reconstructions along the length of the piston may be more accurate in estimating the position of stapes implants.

Entities:  

Mesh:

Year:  2008        PMID: 18520630     DOI: 10.1097/MAO.0b013e3181758e96

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  4 in total

1.  Analysis of the accuracy of high-resolution computed tomography techniques for the measurement of stapes prostheses.

Authors:  Alessandro Bozzato; Tobias Struffert; Victoria Hertel; Heinrich Iro; Joachim Hornung
Journal:  Eur Radiol       Date:  2009-09-02       Impact factor: 5.315

2.  Surgical and clinical confirmation of temporal bone CT findings in patients with otosclerosis with failed stapes surgery.

Authors:  J Whetstone; A Nguyen; A Nguyen-Huynh; B E Hamilton
Journal:  AJNR Am J Neuroradiol       Date:  2014-01-30       Impact factor: 3.825

3.  Imaging Studies in Otosclerosis: An Up-to-date Comprehensive Review.

Authors:  Lucas Resende Lucinda Mangia; Luiz Otávio de Mattos Coelho; Bettina Carvalho; Adriana Kosma Pires de Oliveira; Rogerio Hamerschmidt
Journal:  Int Arch Otorhinolaryngol       Date:  2020-09-24

4.  Revision Stapes Surgery.

Authors:  Hitomi Sakano; Jeffrey P Harris
Journal:  Curr Otorhinolaryngol Rep       Date:  2022-01-24
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.