Literature DB >> 18516397

Assessing Ankle-Brachial Index (ABI) by using automated oscillometric devices.

Takao Kawamura1.   

Abstract

BACKGROUND: Assessing Ankle-Brachial Index is an essential procedure in clinical settings, but since its measurement by the gold standard Doppler Ultrasonic (DU) technique is impaired by technical difficulties, it is underperformed.
OBJECTIVE: The aim of this study was to assess the efficacy of an automated oscillometric device (AOD) by performing Ankle-Brachial Index (ABI) assessments and to suggest delta brachial-brachial (delta-BB) and delta-ABI as markers of cardiovascular risk.
METHODS: In this observational and descriptive study, 247 patients (56.2% females, mean age 62.0 years) had their arterial blood pressure (ABP) measured for ABI calculation. Two AOD (OMRON-HEM705CP) devices were used for simultaneous measurements of the ABP, first of the two arms and then of the arm with higher systolic ABP and a leg, first the left and then the right one. When leg ABP measurements were not possible, ABI determination was performed by using the standard Doppler Ultrasonic (DU) technique. Patients were designated to Group N (normal ABI: 0.91 to 1.30) or Group A (abnormal ABI: < or =0.90 or >1.30). Other indexes were also calculated: delta-BB (absolute difference in mmHg of systolic ABP between arms) and delta-ABI (absolute difference of ABI between legs) and the results were compared.
RESULTS: In most patients (90.7%), it was possible to determine the ABI. Group N data allowed calculation of the 95th percentile reference values (RV) of delta-BB (0 to 8 mmHg) and delta-ABI (0 to 0.13). When compared to Group N, Group A had a significantly higher prevalence of high values greater than the RVs of delta-ABI (30 of 52 and 10 of 195, respectively; Odds Ratio = 25.23; p<0.0001) and delta-BB (13 of 52 and 7 of 195, respectively; Odds Ratio = 8.95; p<0.0001).
CONCLUSION: In most patients, the ABI could be measured by AOD. Both indexes, delta-BB and delta-ABI greater than the RVs, were significantly more prevalent in patients with abnormal ABI values, and their usefulness as new markers of cardiovascular disease should be further appraised in epidemiological studies.

Entities:  

Mesh:

Year:  2008        PMID: 18516397     DOI: 10.1590/s0066-782x2008000500003

Source DB:  PubMed          Journal:  Arq Bras Cardiol        ISSN: 0066-782X            Impact factor:   2.000


  4 in total

Review 1.  Peripheral artery disease (PAD) screening in the asymptomatic population: why, how, and who?

Authors:  Henrik Sillesen; Erling Falk
Journal:  Curr Atheroscler Rep       Date:  2011-10       Impact factor: 5.113

2.  The difference in blood pressure readings between arms and survival: primary care cohort study.

Authors:  Christopher E Clark; Rod S Taylor; Angela C Shore; John L Campbell
Journal:  BMJ       Date:  2012-03-20

3.  Comparison of the ankle-brachial index with parameters of stiffness and peripheral arterial resistance assessed by photoplethysmography in elderly patients.

Authors:  Kelser de Souza Kock; João Batista Ferreira da Silva; Jefferson Luiz Brum Marques
Journal:  J Vasc Bras       Date:  2019-05-15

4.  Assessment of peripheral artery disease risk in building construction workers by Ankle-Brachial index measurement with automated oscillometric and hand-held Doppler device.

Authors:  Manish K Verma; Vibha Gangwar; Rajani B Jasrotia; Nitin A John
Journal:  J Family Med Prim Care       Date:  2022-01-31
  4 in total