| Literature DB >> 18514119 |
Dimitrios S Mastrokalos1, Hans H Paessler.
Abstract
We report a case of a systemic allergic reaction to biodegradable poly-L-lactic acid (PLLA) interference screws after anterior cruciate ligament (ACL) reconstruction with bone-patellar tendon-bone graft. A 30-year-old patient complained of certain symptoms, such as an inability to focus mentally, rash on the right femur, chronic fatigue, decreased sex drive, and localized alopecia, 3 months after ACL reconstruction in the right knee. Two biodegradable PLLA interference screws had been used for proximal and distal graft fixation. Allergy testing showed a value of 7 in PLLA antigen. After removal of 1 screw in August 2000, the patient reported marked improvement, but some symptoms remained. In July 2001 he underwent arthroscopic revision ACL reconstruction with hamstrings via an implant-free technique with intensive debridement of the tunnels and removal of all scar tissue and screw rests. All symptoms disappeared, and the PLLA antigen number fell from 3 to "concentrate" 2 months postoperatively.Entities:
Mesh:
Substances:
Year: 2007 PMID: 18514119 DOI: 10.1016/j.arthro.2006.12.029