Sudha Natarajan1, Daniel G Remick.
Abstract
The enzyme-linked immunosorbent assay (ELISA) is often used to measure protein levels in plasma and other solutions. In order for the assay to be quantitative, a standard curve must be prepared for each assay. Technical blunders in preparing the standard curve can render otherwise representative sample values useless. In order to recover these data, a protocol has been developed whereby a new standard curve is generated using the DeltaOD values of the samples. This new standard is applied to the original plate in order to determine sample concentrations. To validate this method, sample concentrations obtained using an acceptable standard curve were plotted against those obtained using the new standard curve. Linear regression analysis showed a 1:1 correlation between concentrations, with r(2) values ranging from 0.98-0.99 and slopes ranging from 0.97-1.10. In this manner, data for these samples are preserved, saving the investigator the time and money involved in repeating experiments.Mesh:
Substances:
Year: 2008 PMID: 18508074 PMCID: PMC2526547 DOI: 10.1016/j.jim.2008.04.001