Athanasios G Manolis1, Dimitrios G Chatzis, Kostas Kouvelas, Zenon S Kyriakides.
Abstract
We describe a case of partial inhibition of ongoing antitachycardia pacing scheme in an implantable cardioverter defibrillator (ICD) recipient with heart failure of ischemic etiology. The cause of inhibition was related to intermittent oversensing of paced T wave and sensing of intrinsic ventricular activation. In our effort to solve this problem, we reprogrammed the paced ventricular blanking period to a higher level.Entities:
Mesh:
Year: 2008 PMID: 18507556 DOI: 10.1111/j.1540-8159.2008.01087.x