Literature DB >> 18507556

Partial inhibition of ongoing antitachycardia pacing sequence due to T-wave oversensing.

Athanasios G Manolis1, Dimitrios G Chatzis, Kostas Kouvelas, Zenon S Kyriakides.   

Abstract

We describe a case of partial inhibition of ongoing antitachycardia pacing scheme in an implantable cardioverter defibrillator (ICD) recipient with heart failure of ischemic etiology. The cause of inhibition was related to intermittent oversensing of paced T wave and sensing of intrinsic ventricular activation. In our effort to solve this problem, we reprogrammed the paced ventricular blanking period to a higher level.

Entities:  

Mesh:

Year:  2008        PMID: 18507556     DOI: 10.1111/j.1540-8159.2008.01087.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  T wave oversensing.

Authors:  Johnson Francis
Journal:  Indian Pacing Electrophysiol J       Date:  2010-06-05