Does patient position influence Doppler signal quality from the USCOM ultrasonic cardiac output monitor?

Lyndon Siu1, Adam Tucker, Shashi Kanth Manikappa, John Monagle.   

Abstract

BACKGROUND: The USCOM1A continuous wave cardiac output monitor (USCOM Pty Ltd., Sydney, NSW, Australia) is a novel Doppler-based device used to measure cardiac output noninvasively. The proper alignment of the transducer, and hence the ultrasound beam to the aortic or pulmonary outflow tracts, is essential to acquire accurate measurements and often much time is spent on transducer and/or patient positioning. In this prospective, observational, crossover study, we investigated the effect of patient positioning on the acquisition of cardiac output measurement with USCOM1A.
METHODS: We measured cardiac output using USCOM1A in 30 healthy adult volunteers, each in five different positions: sitting, supine, Trendelenburg (20 degrees), left lateral tilt (20 degrees), and right lateral tilt (20 degrees) and compared the time required to obtain acceptable measurements. We also compared the quality of the Doppler signal obtained in these positions using a scoring system designed for this study.
RESULTS: There was a higher rate of failed measurement, the mean time to obtain the first acceptable measurement was prolonged and the optimal measurement obtained within a 5-min period was of a lower quality in the sitting position compared with the other four positions.
CONCLUSIONS: Our results suggested the sitting position is the least suitable and least reliable position in which to perform cardiac output measurements using USCOM1A compared with the supine, Trendelenburg (20 degrees), left lateral tilt (20 degrees), and right lateral tilt (20 degrees) positions.

Entities:  

Mesh:

Year:  2008        PMID: 18499612     DOI: 10.1213/ane.0b013e3181732127

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  2 in total

1.  Cardiac index measurements by transcutaneous Doppler ultrasound and transthoracic echocardiography in adult and pediatric emergency patients.

Authors:  H Bryant Nguyen; Daryl P Banta; Gail Stewart; Tommy Kim; Ramesh Bansal; James Anholm; William A Wittlake; Stephen W Corbett
Journal:  J Clin Monit Comput       Date:  2010-06-20       Impact factor: 2.502

2.  Influence of different positions on hemodynamics derived from noninvasive transcutaneous Doppler ultrasound.

Authors:  Cangel Pui-Yee Chan; Pui-Ling Cheung; Mandy Man Tse; Nandini Agarwal; Sangeeta Narain; Stewart Siu-Wa Chan; Brendan E Smith; Colin A Graham; Timothy H Rainer
Journal:  Physiol Rep       Date:  2013-09-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.