The impact of MLC transmitted radiation on EPID dosimetry for dynamic MLC beams.

Philip Vial1, Peter B Greer, Peter Hunt, Lyn Oliver, Clive Baldock.   

Abstract

The purpose of this study was to experimentally quantify the change in response of an amorphous silicon (a-Si) electronic portal imaging device (EPID) to dynamic multileaf collimator (dMLC) beams with varying MLC-transmitted dose components and incorporate the response into a commercial treatment planning system (TPS) EPID prediction model. A combination of uniform intensity dMLC beams and static beams were designed to quantify the effect of MLC transmission on EPID response at the central axis of 10 x 10 cm2 beams, at off-axis positions using wide dMLC beam profiles, and at different field sizes. The EPID response to MLC transmitted radiation was 0.79 +/- 0.02 of the response to open beam radiation at the central axis of a 10 x 10 cm2 field. The EPID response to MLC transmitted radiation was further reduced relative to the open beam response with off-axis distance. The EPID response was more sensitive to field size changes for MLC transmitted radiation compared to open beam radiation by a factor of up to 1.17 at large field sizes. The results were used to create EPID response correction factors as a function of the fraction of MLC transmitted radiation, off-axis distance, and field size. Software was developed to apply the correction factors to each pixel in the TPS predicted EPID image. The corrected images agreed more closely with the measured EPID images in areas of intensity modulated fields with a large fraction of MLC transmission and, as a result the accuracy of portal dosimetry with a-Si EPIDs can be improved. Further investigation into the detector response function and the radiation source model are required to achieve improvements in accuracy for the general case.

Entities:  

Mesh:

Year:  2008        PMID: 18491519     DOI: 10.1118/1.2885368

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy.

Authors:  Aiping Ding; Lei Xing; Bin Han
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

2.  Predicting gamma evaluation results of patient-specific head and neck volumetric-modulated arc therapy quality assurance based on multileaf collimator patterns and fluence map features: A feasibility study.

Authors:  Sangutid Thongsawad; Somyot Srisatit; Todsaporn Fuangrod
Journal:  J Appl Clin Med Phys       Date:  2022-05-18       Impact factor: 2.243

3.  Evaluation of a fast method of EPID-based dosimetry for intensity-modulated radiation therapy.

Authors:  Benjamin E Nelms; Karl H Rasmussen; Wolfgang A Tome
Journal:  J Appl Clin Med Phys       Date:  2010-04-19       Impact factor: 2.102

4.  A fully electronic intensity-modulated radiation therapy quality assurance (IMRT QA) process implemented in a network comprised of independent treatment planning, record and verify, and delivery systems.

Authors:  Daniel W Bailey; Lalith Kumaraswamy; Matthew B Podgorsak
Journal:  Radiol Oncol       Date:  2010-05-24       Impact factor: 2.991

5.  aSi-EPID transit signal calibration for dynamic beams: a needful step for the IMRT in vivo dosimetry.

Authors:  Francesca Greco; Angelo Piermattei; Luigi Azario; Lorenzo Placidi; Savino Cilla; Rocchina Caivano; Vincenzo Fusco; Andrea Fidanzio
Journal:  Med Biol Eng Comput       Date:  2013-07-09       Impact factor: 2.602

6.  EPID dosimetry for pretreatment quality assurance with two commercial systems.

Authors:  Daniel W Bailey; Lalith Kumaraswamy; Mohammad Bakhtiari; Harish K Malhotra; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2012-07-05       Impact factor: 2.102

7.  Intensity-modulated radiation therapy dose verification using fluence and portal imaging device.

Authors:  Iori Sumida; Hajime Yamaguchi; Indra J Das; Hisao Kizaki; Keiko Aboshi; Mari Tsujii; Yuji Yamada; Osamu Suzuki; Yuji Seo; Fumiaki Isohashi; Kazuhiko Ogawa
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

8.  Calculation of exit dose for conformal and dynamically-wedged fields, based on water-equivalent path length measured with an amorphous silicon electronic portal imaging device.

Authors:  Awusi Kavuma; Martin Glegg; Mohamed Metwaly; Garry Currie; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2011-03-03       Impact factor: 2.102

9.  Optimized Varian aSi portal dosimetry: development of datasets for collective use.

Authors:  Ann Van Esch; Dominique P Huyskens; Lukas Hirschi; Christof Baltes
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

10.  Pixel response-based EPID dosimetry for patient specific QA.

Authors:  Bin Han; Aiping Ding; Minghui Lu; Lei Xing
Journal:  J Appl Clin Med Phys       Date:  2016-12-15       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.