| Literature DB >> 18490203 |
Philippe A Guy1, Isabelle Tavazzi, Stephen J Bruce, Ziad Ramadan, Sunil Kochhar.
Abstract
Optimisation and method validation was assessed here for metabolic profiling analysis of urine samples using UPLC-TOFMS. A longer run time of 31 min revealed greater reproducibility, and the higher number of variables was identified as compared to shortened run times (10 and 26 min). We have also implemented two QC urine samples enabling the assessment of the quality and reproducibility of the data generated during the whole analytical workflow (retention time drift, mass precision and fluctuation of the ion responses over time). Based on the QC data, suitable standards for ensuring consistent analytical results for metabolomics applications using the UPLC-MS techniques are recommended.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18490203 DOI: 10.1016/j.jchromb.2008.04.034
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205