Use of near-infrared spectroscopy as a physiologic monitor for intra-abdominal hypertension.

Sandy Widder1, Michael K Ranson, David Zygun, Linda Knox, Kevin B Laupland, Philip Laird, Chad G Ball, Andrew W Kirkpatrick.   

Abstract

BACKGROUND: Intra-abdominal hypertension (IAH) is increasingly recognized as an important parameter in critically ill (ICU) patients. IAH affects perfusion to all abdominal components including the abdominal wall (AW). Near infrared spectroscopy (NIRS) measures changes in three chromophores including oxygenated hemoglobin (HbO2), deoxygenated hemoglobin (Hb), and cytochrome aa3 (Cyt), providing information concerning dysoxia. We sought to examine whether NIRS measurement of the AW was safe, and correlated with intra-abdominal pressure (IAP) in ICU patients.
METHODS: A NIU-Pro001 system recorded NIRS data over 24 hours from the AW of 9 ICU patients at risk for IAH. IAP was recorded from the bladder. Changes in chromophore values are interpolated from best-fit polynomial curves.
RESULTS: Sixty-six paired IAP and NIRS readings were taken from 9 ICU (4-12 observations/patient) patients. No measurement related adverse reactions occurred. The mean (range) first values measured in these patients were; IAP 17.2 mm Hg (9-31); HbO2 0.41 micromol/L (-8.4 to 7.6); Hb 2.6 micromol/L (-3.1 to 12.2); and Cyt 0.65 micromol/L (-3.4 to 4.8). A significant, inverse (or negative) association was found between DeltaNIRS HbO2 level and DeltaIAP (coefficient, -0.1588; p = 0.008). No association was found between DeltaNIRS Hb or DeltaCyt and DeltaIAP.
CONCLUSION: NIRS of the AW appears safe. NIRS measurement of changes in HbO2 appears to be associated with DeltaIAP, and warrant further study in greater numbers of ICU patients with more frequent IAP readings, over longer periods of critical illness.

Entities:  

Mesh:

Year:  2008        PMID: 18469636     DOI: 10.1097/TA.0b013e31814695dd

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  2 in total

1.  Splanchnic-cerebral oxygenation ratio (SCOR) values in healthy term infants as measured by near-infrared spectroscopy (NIRS).

Authors:  Sean M Bailey; Karen D Hendricks-Muñoz; Pradeep Mally
Journal:  Pediatr Surg Int       Date:  2013-02-28       Impact factor: 1.827

2.  Somatic regional oxygen saturation as an early marker of intra-abdominal hypertension in critically ill children: a pilot study

Authors:  Özden Özgür Horoz; Nagehan Aslan; Dinçer Yildizdaş; Yasemin Çoban; Yaşar Sertdemir; Awni Al-Subu
Journal:  Turk J Med Sci       Date:  2020-04-09       Impact factor: 0.973

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.