| Literature DB >> 18465745 |
Daniel M Goldenholz1, Seppo P Ahlfors, Matti S Hämäläinen, Dahlia Sharon, Mamiko Ishitobi, Lucia M Vaina, Steven M Stufflebeam.
Abstract
Although magnetoencephalography (MEG) and electroencephalography (EEG) have been available for decades, their relative merits are still debated. We examined regional differences in signal-to-noise-ratios (SNRs) of cortical sources in MEG and EEG. Data from four subjects were used to simulate focal and extended sources located on the cortical surface reconstructed from high-resolution magnetic resonance images. The SNR maps for MEG and EEG were found to be complementary. The SNR of deep sources was larger in EEG than in MEG, whereas the opposite was typically the case for superficial sources. Overall, the SNR maps were more uniform for EEG than for MEG. When using a noise model based on uniformly distributed random sources on the cortex, the SNR in MEG was found to be underestimated, compared with the maps obtained with noise estimated from actual recorded MEG and EEG data. With extended sources, the total area of cortex in which the SNR was higher in EEG than in MEG was larger than with focal sources. Clinically, SNR maps in a patient explained differential sensitivity of MEG and EEG in detecting epileptic activity. Our results emphasize the benefits of recording MEG and EEG simultaneously. 2008 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2009 PMID: 18465745 PMCID: PMC2882168 DOI: 10.1002/hbm.20571
Source DB: PubMed Journal: Hum Brain Mapp ISSN: 1065-9471 Impact factor: 5.038