Literature DB >> 18463521

Analyses of cutting and aspirating properties of vitreous cutters with high-speed camera.

Tatsuhiko Sato1, Shunji Kusaka, Yusuke Oshima, Takashi Fujikado.   

Abstract

PURPOSE: To investigate the cutting and aspirating properties of eight vitreous cutters under different conditions and to analyze the factors determining these properties.
METHODS: Commercially available 20-gauge (20-G), 23-G, and 25-G pneumatic vitreous cutters were evaluated. The areas of the inner tube and cutting port were measured with a microscope and 3D-CAD software. The total opening area/s of the cutting port was determined by photographing the movements of the cutter blade with a high-speed camera. The total volume of balanced salt solution aspirated/s was measured.
RESULTS: The inner tube areas ranged from 0.350-0.352 mm2 for the 20-G, 0.169-0.196 mm for the 23-G, and 0.128-0.129 mm2 for the 25-G cutters. The port areas ranged from 0.254 to 0.306 mm2 for the 20-G, 0.122 to 0.173 mm2 for the 23-G, and 0.066 to 0.125 mm2 for the 25-G cutters. High-speed camera analyses revealed that the total opening area was determined not only by the port area but also by the duty cycle, and that the latter varied among the cutters even under the same cutting speed. There was a positive correlation (R2 >0.88) between total opening area/s and total aspirated volume/s.
CONCLUSIONS: Larger cutting port areas and higher duty cycles are associated with higher aspirated volumes of vitreous cutters.

Entities:  

Mesh:

Year:  2008        PMID: 18463521     DOI: 10.1097/IAE.0b013e3181631907

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  7 in total

1.  Retrospective comparison of 25-gauge vitrectomy for repair of proliferative vitreoretinopathy with or without anterior proliferation.

Authors:  Tatsuhiko Sato; Kazuyuki Emi; Hajime Bando; Toshihide Ikeda
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-11-07       Impact factor: 3.117

2.  Analysis of a 23-gauge ultra high-speed cutter with duty cycle control.

Authors:  Bruno Diniz; Rodrigo B Fernandes; Ramiro M Ribeiro; Jaw-Chyng Lue; Anderson G Teixeira; Octaviano Magalhães; Mauricio Maia; Mark S Humayun
Journal:  Retina       Date:  2013-05       Impact factor: 4.256

3.  Fluidics in a dual pneumatic ultra high-speed vitreous cutter system.

Authors:  Bruno Diniz; Ramiro M Ribeiro; Rodrigo B Fernandes; Jaw-Chyng Lue; Anderson G Teixeira; Mauricio Maia; Mark S Humayun
Journal:  Ophthalmologica       Date:  2012-10-27       Impact factor: 3.250

4.  Segmentation and removal of fibrovascular membranes with high-speed 23 G transconjunctival sutureless vitrectomy, in severe proliferative diabetic retinopathy.

Authors:  Erkan Celik; Ozkan Sever; Fatih Horozoglu; Ates Yanyalı
Journal:  Clin Ophthalmol       Date:  2016-05-17

5.  Faster recovery after 25-gauge microincision vitrectomy surgery than after 20-gauge vitrectomy in patients with proliferative diabetic retinopathy.

Authors:  Tatsuhiko Sato; Kazuyuki Emi; Hajime Bando; Toshihide Ikeda
Journal:  Clin Ophthalmol       Date:  2012-11-21

6.  Hydraulic Resistance of Vitreous Cutters: The Impact of Blade Design and Cut Rate.

Authors:  Tommaso Rossi; Giorgio Querzoli; Giampiero Angelini; Carlo Malvasi; Alessandro Rossi; Mario Morini; Graziana Esposito; Alessandra Micera; Natale Mario di Luca; Guido Ripandelli
Journal:  Transl Vis Sci Technol       Date:  2016-07-01       Impact factor: 3.283

7.  FIRST-IN-HUMAN CLINICAL STUDY TO INVESTIGATE THE EFFECTIVENESS AND SAFETY OF PARS PLANA VITRECTOMY SURGERY USING A NEW HYPERSONIC TECHNOLOGY.

Authors:  Paulo E Stanga; Jon I Williams; Sherif A Shaarawy; Amar Agarwal; Anusha Venkataraman; Dhivya A Kumar; Timothy T You; Richard S Hope
Journal:  Retina       Date:  2020-01       Impact factor: 3.975

  7 in total