Literature DB >> 18460847

Reduction of a high-field dielectric artifact with homemade gel.

Yukihisa Takayama1, Hiroi Nonaka, Manabu Nakajima, Takayuki Obata, Hiroo Ikehira.   

Abstract

We developed an original dielectric-equivalent gel (Japanese Patent Application ID: P2004-236876A, February 6, 2003) pad that could improve inhomogeneous images caused primarily by B(1) homogeneity on a high-tesla magnetic resonance (MR) system. Made easily from various ingredients, our gel could reduce interference from radiofrequency waves at an object's surface and show changes in B(1) inhomogeneity. We herein assess the gel's effect using a plastic-bottle phantom on a 3T MR system.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18460847     DOI: 10.2463/mrms.7.37

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  1 in total

1.  Permittivity and performance of dielectric pads with sintered ceramic beads in MRI: early experiments and simulations at 3 T.

Authors:  Wei Luo; Michael T Lanagan; Christopher T Sica; Yeunchul Ryu; Sukhoon Oh; Matthew Ketterman; Qing X Yang; Christopher M Collins
Journal:  Magn Reson Med       Date:  2012-08-13       Impact factor: 4.668