| Literature DB >> 18457651 |
Natalie Chavane1, Renaud Jacquemart, Caroline D Hoemann, Mario Jolicoeur, Gregory De Crescenzo.
Abstract
We report an innovative at-line method to monitor concentration of bioactive antibody (i.e., antibody with conserved antigen-binding activity) secreted during bioreactor culture by the use of surface plasmon resonance (SPR)-based biosensor technology. In a first series of experiments, conditions for SPR-based measurements were validated off-line by monitoring bioactive antibody concentration in conditioned medium from 500-ml baffled flask hybridoma cell cultures. A fully automated experimental setup in which the SPR-based biosensor was harnessed to a bioreactor was then used at-line to monitor the concentration of bioactive antibody produced in a 3.5-L bioreactor. Quantitative SPR measurements performed both at-line and off-line were in excellent agreement with quantitative Western blotting followed by densitometry analyses. Thus, our experimental study confirms that SPR biosensors can be applied to at-line quantification of correctly folded proteins that are secreted by cells cultured in a bioreactor. Our experimental approach represents a novel and robust analytical strategy to be applied to the control and optimization of the production of bioactive secreted proteins.Mesh:
Substances:
Year: 2008 PMID: 18457651 DOI: 10.1016/j.ab.2008.04.019
Source DB: PubMed Journal: Anal Biochem ISSN: 0003-2697 Impact factor: 3.365