Agreement and repeatability of the Sonogage ultrasound pachometer compared with a Mitutoyo micrometer.

James W Weitkamp1, Harue J Marsden, William Berke, Grace Daijo.   

Abstract

PURPOSE: The agreement and repeatability of the Sonogage ultrasound pachometer was tested by comparing it with a Mitutoyo micrometer.
METHODS: The thickness of five polymacon disks with thickness and ultrasonic properties similar to the human cornea and corneal epithelium were measured using both instruments. The mean values obtained using both instruments were compared and analyzed using a two-way analysis of variance, Bland-Altman analysis, and coefficient of variation.
RESULTS: The measurements of the Sonogage were found to be statistically indistinguishable from the micrometer for all disks combined (p = 0.25) and for the disks with thicknesses similar to the corneal epithelium (p = 0.10), Bland-Altman analysis reveal limits of agreement for all thicknesses with a lower limit of -4.23 microm and upper limit of 5.27 microm. The coefficient of variation was <4% for each thickness. Bland-Altman plots of thin and thick disks exhibit no systematic difference between the average micrometer and Sonogage measurements.
CONCLUSION: The measurements taken with the Sonogage pachometer are comparable with those taken with the micrometer when measuring polymacon disks with properties similar to corneal epithelium and total corneal thickness.

Entities:  

Mesh:

Year:  2008        PMID: 18451740     DOI: 10.1097/OPX.0b013e31816bf55d

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  1 in total

1.  Genetic dependence of central corneal thickness among inbred strains of mice.

Authors:  Geoffrey D Lively; Bing Jiang; Adam Hedberg-Buenz; Bo Chang; Greg E Petersen; Kai Wang; Markus H Kuehn; Michael G Anderson
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-08-26       Impact factor: 4.799

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.