Biomechanical comparison of a 2 and 3 proximal screw-configured antegrade piriformis intramedullary nail with a trochanteric reconstruction nail in an unstable subtrochanteric fracture model.

Brian Fissel1, Berton R Moed, J Gary Bledsoe.   

Abstract

OBJECTIVE: The objective of this study was to test the stiffness and ultimate load to failure of new intramedullary (IM) nail proximal screw configurations as compared to a trochanteric reconstruction nail.
METHODS: Twenty-one synthetic composite femurs were mounted on a Material Testing System and tested in axial compression 5 times. The femurs had an 1 of 2 IM nail types inserted with 1 of 3 proximal screw configurations: a 3-screw configuration with 2 transverse screws and a screw angled into the femoral neck; a 2-screw design with a single transverse screw and a single screw angled into the femoral neck; 2 parallel screws angled into the femoral neck. There were 7 specimens in each group. An unstable fracture (OTA/AO 32-C3.2) was created. and the stiffness of these constructs was tested in compression 5 times. Each construct was then loaded to failure in compression.
RESULTS: The 3-screw construct provided more axial stiffness (214 N/mm +/- 75) than either the 2-screw construct (123 N/mm +/- 32) or the trochanteric reconstruction nail (127 N/mm +/- 21) (P = 0.017 and 0.035 for 3-screw vs. 2-screw and recon respectively, P = 0.45 for 2-screw vs. recon). Load-to-failure testing demonstrated similarity among the different screw configurations (3-screw = 2230 N +/- 265, 2-screw = 2283 N +/- 260, Reconstruction nail = 2121 N +/- 156) (P = 1.0 all groups).
CONCLUSIONS: The proximal 3-screw configuration provided more stiffness than either the 2-screw configuration or trochanteric reconstruction nail. The 2-screw configuration performed equally to a standard trochanteric reconstruction nail in stiffness testing. The ultimate loads to failure for the 3 tested constructs were not significantly different.

Entities:  

Mesh:

Year:  2008        PMID: 18448988     DOI: 10.1097/BOT.0b013e318172c7f4

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

1.  Four pins assisted reduction of complex segmental femoral fractures: a technique for closed reduction.

Authors:  Zhan-le Zheng; Xian Yu; Guo-Qiang Xu; Wei Chen; Ying-Ze Zhang; Zhen-Qing Jiao
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2014-12-06

2.  Comparison of standard versus reconstruction proximal interlocking screw configurations for antegrade intramedullary nail fixation of femoral shaft fractures.

Authors:  Jihyo Hwang; Darryl F Cannady; Samantha Nino; Kenneth J Koval; Joshua R Langford; Joshua A Parry
Journal:  J Clin Orthop Trauma       Date:  2021-03-01

3.  Biomechanical Evaluation of Four Methods for Internal Fixation of Comminuted Subtrochanteric Fractures.

Authors:  Jie Wang; Jian-Xiong Ma; Hao-Bo Jia; Yang Chen; Yang Yang; Xin-Long Ma
Journal:  Medicine (Baltimore)       Date:  2016-05       Impact factor: 1.889

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.