Literature DB >> 18448424

Patient- and lead-related factors affecting lead fracture in children with transvenous permanent pacemaker.

Hasim Olgun1, Tevfik Karagoz, Alpay Celiker, Naci Ceviz.   

Abstract

AIMS: Fracture in transvenous pacing leads is one of the most common reasons for lead abandonment. Although the factors affecting lead failure rates have been investigated, there is no study evaluating the clinical parameters that affect lead fracture in children. We report our experience with lead fracture in children with transvenous pacemakers. METHODS AND
RESULTS: The follow-up results of 264 leads from 184 patients were evaluated using pacemaker follow-up data. Underlying conditions, implant data, and lead features were evaluated for the analysis of lead fracture. During a mean follow-up of 72.8 +/- 39.7 months (range 3.2-160.6, median 70), lead fracture developed in 19 leads (7.2%) from 18 patients. The mean duration between implantation and lead fracture was 57.3 +/- 35 months (range 6.8-130, median 51). All fractures occurred in the leads implanted by the infraclavicular subclavian approach. Cumulative survival at the end of 5 years was 92.7% in terms of lead fracture. None of the patient-related risk factors correlated with lead fracture. Multivariate analyses of lead-related risk factors revealed a significant correlation only between lead fracture and fixation mechanism (P < 0.05).
CONCLUSION: Our results indicated that none of the patient-related risk factors was correlated with lead fracture. Among lead-related risk factors, only the fixation mechanism was found to be correlated with lead fracture; thus, it seems that passive fixation mechanism is safer in terms of lead fracture. Although all fractures occurred in the leads implanted by the intrathoracic subclavian approach, statistical analysis revealed no significance for this parameter. The effect of the extrathoracic approach should be investigated in a large group of patients.

Entities:  

Mesh:

Year:  2008        PMID: 18448424     DOI: 10.1093/europace/eun109

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  4 in total

1.  Distal pacemaker lead fracture: a rare entity.

Authors:  D X Augustine; K Carson; A Garg
Journal:  BMJ Case Rep       Date:  2010-09-29

2.  Successful implantation of leadless pacemakers in children: a case series.

Authors:  Ewa Jędrzejczyk-Patej; Aleksandra Woźniak; Linda Litwin; Alina Skiba-Zdrzałek; Michał Mazurek; Radosław Lenarczyk; Zbigniew Kalarus; Oskar Kowalski
Journal:  Eur Heart J Case Rep       Date:  2020-03-30

3.  Long-term reliability of the defibrillator lead inserted by the extrathoracic subclavian puncture.

Authors:  Masaya Watanabe; Hisashi Yokoshiki; Hirofumi Mitsuyama; Kazuya Mizukami; Taro Tenma; Rui Kamada; Masayuki Takahashi; Yoshiro Matsui; Toshihisa Anzai
Journal:  J Arrhythm       Date:  2018-08-18

4.  Multiple pacemaker lead breakages due to clavicle dislocation following clavicle fracture.

Authors:  Hideyuki Hasebe
Journal:  Indian Pacing Electrophysiol J       Date:  2017-07-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.