Precision sampling measurements using ac programmable Josephson voltage standards.

A Rüfenacht1, C J Burroughs, S P Benz.   

Abstract

We have performed a variety of precision measurements by comparing ac and dc waveforms generated by two independent ac programmable Josephson voltage standard (ACPJVS) systems. The objective of these experiments was to demonstrate the effectiveness of using a sampling digital voltmeter to measure small differences between Josephson waveforms for frequencies up to 3.6 kHz. The low uncertainties that we obtained confirm the feasibility of using this differential sampling method for high accuracy comparisons between ACPJVS waveforms and signals from other sources.

Year:  2008        PMID: 18447541     DOI: 10.1063/1.2901683

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Uncertainty analysis for ac-dc difference measurements with the AC Josephson Voltage Standard.

Authors:  Jason M Underwood
Journal:  Metrologia       Date:  2018       Impact factor: 3.157