Literature DB >> 18447518

Simple active method for reducing magnetic interference in a thermoelectrically cooled photomultiplier tube.

Petri Kärhä1, Teemu Jaakkola.   

Abstract

We present a simple modification for thermoelectrically cooled photomultiplier tube (PMT) assemblies that eliminates the magnetic interference between the peltier element and the PMT. An active compensation is accomplished by forming current loops of the wires of the peltier element and placing them in such a way that they eliminate the interfering magnetic field. It is demonstrated that the improved system reduces measurement errors of the order of 1% to statistical noise at the level of 0.07%.

Year:  2008        PMID: 18447518     DOI: 10.1063/1.2906317

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  OpenTCC: An open source low-cost temperature-control chamber.