Literature DB >> 18446554

Autoclaving and battery capacity influence on laser fluorescence measurements.

Renata Maciel Rocha-Cabral1, Fausto Medeiros Mendes, Fabiana Miura, Adriana Da Costa Ribeiro, Mariana Minatel Braga, Denise Maria Zezell.   

Abstract

AIM: To evaluate the influence of probe tip autoclaving and depleted alkaline batteries on laser fluorescence (LF-DIAGNOdent) device performance.
MATERIAL AND METHODS: One-hundred-and-twenty occlusal sites were analyzed with an LF device in nine conditions: No autoclaved probe and probe after 5, 10, 15, 20, 25, 30, 40, and 50 autoclaving cycles. Subsequently, the performance of the device was analyzed with: New batteries, batteries with 1.49/1.39 V, 1.38/1.37 V, 1.36/1.34 V, 1.33/1.32 V, and lower than 1.32 V. LF values, sensitivity, specificity, accuracy, and area under the ROC curve were compared.
RESULTS: In the probe study, sensitivity was lower after 50 sterilizing cycles, though specificity was higher than the assessment performed using a new tip. In the batteries study, specificity was higher for depleted batteries, but LF performance did not differ significantly among the groups.
CONCLUSION: Batteries do not significantly influence device performance, but consecutive sterilization of probes in autoclave alters readings, downgrading its performance.

Mesh:

Year:  2008        PMID: 18446554     DOI: 10.1080/00016350802017484

Source DB:  PubMed          Journal:  Acta Odontol Scand        ISSN: 0001-6357            Impact factor:   2.331


  2 in total

1.  Influence of cross-infection control methods on performance of pen-type laser fluorescence in detecting occlusal caries lesions in primary teeth.

Authors:  Ronilza Matos; Tatiane Fernandes Novaes; Alessandra Reyes; Monique Saveriano De Benedetto; Fausto Medeiros Mendes; Mariana Minatel Braga
Journal:  Lasers Med Sci       Date:  2012-05-09       Impact factor: 3.161

2.  Fluorescence devices for the detection of dental caries.

Authors:  Richard Macey; Tanya Walsh; Philip Riley; Anne-Marie Glenny; Helen V Worthington; Patrick A Fee; Janet E Clarkson; David Ricketts
Journal:  Cochrane Database Syst Rev       Date:  2020-12-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.