Comparision between bed side testing of blood glucose by glucometer vs centralized testing in a tertiary care hospital.

Ayaz Baig1, Imran Siddiqui, Abdul Jabbar, Syed Iqbal Azam, Salman Sabir, Shahryar Alam, Farooq Ghani.   

Abstract

BACKGROUND: To determine the accuracy, turnaround time and cost effectiveness of bedside monitoring of blood glucose levels by non-laboratory health care workers and centralized testing of blood glucose by automated analyzer in a tertiary care hospital.
METHODS: The study was conducted in Section of Chemical Pathology, Department of Pathology and Microbiology and Section of Endocrinology Department of Medicine, Aga Khan University and Hospital Karachi, from April 2005 to March 2006. One hundred and ten patients were included in the study. The blood glucose levels were analyzed on glucometer (Precision Abbott) by finger stick, using Biosensor Technology. At the same time venous blood was obtained to analyze glucose in clinical laboratory on automated analyzer (SYNCHRON CX7) by glucose oxidase method.
RESULTS: We observed good correlation between bed side glucometer and laboratory automated analyzer for glucose values between 3.3 mmol/L (60 mg/dl) and 16.7 (300 mg/dl). A significant difference was observed for glucose values less than 3.3 mmol/L (p = 0.002) and glucose values more than 16.67 mmol/l (p = 0.049). Mean Turnaround time for glucometer and automated analyzer were 0.08 hours and 2.49 hours respectively. The cost of glucose testing with glucometer was 48.8% lower than centralized lab based testing.
CONCLUSION: Bedside glucometer testing, though less expensive does not have good accuracy in acutely ill patient with either very high or very low blood glucose levels.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18444586

Source DB:  PubMed          Journal:  J Ayub Med Coll Abbottabad        ISSN: 1025-9589


  4 in total

Review 1.  Social, organizational, and contextual characteristics of clinical decision support systems for intensive insulin therapy: a literature review and case study.

Authors:  Thomas R Campion; Lemuel R Waitman; Addison K May; Asli Ozdas; Nancy M Lorenzi; Cynthia S Gadd
Journal:  Int J Med Inform       Date:  2009-10-07       Impact factor: 4.046

Review 2.  Blood glucose measurement in the intensive care unit: what is the best method?

Authors:  Huong T Le; Neil S Harris; Abby J Estilong; Arvid Olson; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

3.  A comparative assessment of the glucose monitor (SD Codefree) and auto analyzer (BT-3000) in measuring blood glucose concentration among diabetic patients.

Authors:  Osman Nabayire Kanwugu; Gideon Kofi Helegbe; Paul Armah Aryee; Nathaniel Amigamsa Akontatiba; Jacob Ankrah; Nsoh Godwin Anabire; Frank Anaba; Benjamin Ahenkora
Journal:  BMC Res Notes       Date:  2017-09-06

4.  Is serum zinc level associated with prediabetes and diabetes?: a cross-sectional study from Bangladesh.

Authors:  Md Rafiqul Islam; Iqbal Arslan; John Attia; Mark McEvoy; Patrick McElduff; Ariful Basher; Waliur Rahman; Roseanne Peel; Ayesha Akhter; Shahnaz Akter; Khanrin P Vashum; Abul Hasnat Milton
Journal:  PLoS One       Date:  2013-04-17       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.