Literature DB >> 18439603

Evaluation of a disposable plastic Neubauer counting chamber for semen analysis.

Jackson Kirkman-Brown1, Lars Björndahl.   

Abstract

OBJECTIVE: To evaluate whether disposable plastic counting chambers effectively could replace nondisposable, time-consuming, and potentially dangerous glass hemocytometers.
DESIGN: Evaluation of equipment in modern laboratory andrology. Comparison of results obtained with plastic chambers with results obtained with "gold-standard" glass hemocytometer counts.
SETTING: Diagnostic laboratory for andrology. PATIENT(S): Twenty-one patients undergoing investigation for infertility problems. INTERVENTION(S): No interventions with patients; sperm in diluted semen samples were used when patients had allowed the use for research and training. MAIN OUTCOME MEASURE(S): Sperm concentration, difference from results obtained with standard equipment. RESULT(S): In the first three experimental series, with use of standard routine phase-contrast microscopy, significantly lower count results were obtained consistently from the plastic chambers than from standard chambers. In the fourth series, with use of specialized equipment, equivalent results were obtained but with a considerably greater time commitment because of difficulties in distinguishing sperm adjacent to the gridlines in the plastic chambers. CONCLUSION(S): The plastic disposable chamber type was not suitable for routine semen analysis because results are variable depending on the microscope used, and increased time is necessary to do the assessment accurately.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18439603     DOI: 10.1016/j.fertnstert.2007.11.076

Source DB:  PubMed          Journal:  Fertil Steril        ISSN: 0015-0282            Impact factor:   7.329


  2 in total

1.  Rmax: A systematic approach to evaluate instrument sort performance using center stream catch.

Authors:  Andrew Riddell; Rui Gardner; Alexis Perez-Gonzalez; Telma Lopes; Lola Martinez
Journal:  Methods       Date:  2015-03-04       Impact factor: 3.608

2.  Improvement of cell counting method for Neubauer counting chamber.

Authors:  Minghao Zhang; Lingui Gu; Peihua Zheng; Zhixin Chen; Xinqi Dou; Qizhong Qin; Xiaozhong Cai
Journal:  J Clin Lab Anal       Date:  2019-08-30       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.